COD Digester Manufacturer,Supplier and Exporter in India
Product Code : EL-CD-2303
Jlab Export is a leading COD Digester Manufacturer,and suppliers in India
Chemical oxygen demand shows how much oxidisable matter a water sample holds, and this thermoreactor heats the samples for that test. Its polished aluminium block takes 15 glass tubes fitted with air condensers and heats them up to 150 °C under a microprocessor controller with an RTD sensor. A digital timer starts on its own when the set temperature is reached.
Specifications
| Controller | Microprocessor digital temperature indicator-cum-controller, LED display of set and process values |
| Temperature range | Ambient +5 °C to 150 °C |
| Sensor | RTD (Pt-100) |
| Block | Polished aluminium, 270 x 150 x 70 mm, heater load 1.0 kW |
| Cavities | 15, each 40 mm dia x 65 mm deep |
| Tubes supplied | 15 glass tubes, 38 mm dia x 200 mm long |
| Air condensers supplied | 15, each 600 mm long |
| Sample size | 20 ml |
| Timer | Automatic digital minute timer, starts at set temperature, switches heaters off, buzzer at cycle end |
| Body and power | Powder-coated CRC steel, one tube rack, 230 V AC, 50/60 Hz |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
COD Practicals in Environmental Labs
Students take 20 ml of each well-mixed water sample in a tube, add the dichromate digestion reagent and fit the air condenser. A blank of distilled water goes through the same steps. The block is set to the method temperature, and the timer runs the digestion and switches the heaters off at the end.
After cooling, the dichromate left in each tube is titrated with Ferrous Ammonium Sulphate from a Burette, using ferroin as the indicator. The difference from the blank gives the COD. Many classes compare COD with BOD on the same samples, with the BOD bottles kept in a BOD Incubator with Shaker.
Frequently Asked Questions
Why is a blank run with every batch?
The blank is distilled water treated exactly like the samples. It shows how much dichromate the reagents and glassware use up on their own, and the COD of each sample is worked out from the difference between its titration and the blank titration.
When does the timer start?
The digital minute timer starts by itself once the block reaches the set temperature, so warm-up time is not counted. At the end of the set period the heaters switch off and a buzzer marks the end of the cycle.
What are the air condensers for?
Each 600 mm air condenser sits on top of its tube during digestion. Vapour from the hot sample cools inside it and runs back down, so volatile organic matter stays in the tube and is oxidised instead of escaping.
How do COD and BOD differ?
COD is the oxygen equivalent of the matter a strong chemical oxidant can oxidise, and the test is finished within hours. BOD is the oxygen used by microbes over several days. COD usually reads higher, because the oxidant attacks matter that microbes cannot break down.
What mistakes do students make in the COD test?
Common ones are not mixing the sample before taking it, leaving off a condenser so a tube loses liquid, and titrating before the tubes have cooled. Samples rich in chloride read falsely high unless the chloride correction in the method is used.
Other Equipment for the Water Lab
For a quote, use Add To Cart or send us a message through the contact page.