Microprocessor K.F. Moisture Titrator

Product Code : EL-MT-2207
Jlab Export is a leading Microprocessor K.F. Moisture Titrator Manufacturer,and suppliers in India

This volumetric Karl Fischer titrator finds the water in a sample and shows it directly as ppm, per cent or mg H2O. A motorized dispensing system meters the reagent in 0.01 ml steps, so there is no glass burette or plunger to break. It reads from 10 ppm to 100% moisture and keeps the last 20 results with date, time and batch number.

Technical Details

Measuring methodKarl Fischer volumetric titration
Measuring range10 ppm to 100% moisture
Titrant volume0.00 to 99.99 ml on display, 0.01 ml resolution
Results shown as% (0.00 to 100), ppm (10 to 9999999), mg H2O (0.00 to 999.99)
End pointVisual LED and audible alarm, wait time adjustable 1 to 59 seconds
MemoryLatest 20 samples (FIFO), battery backup for at least 1 month
Display and keys8-digit red seven-segment display, 24 soft-touch membrane keys
StirrerBuilt-in magnetic stirrer, adjustable speed
Reagent bottles3 x 250 ml (KF reagent, methanol, calcium chloride)
Power and size230 V AC, 50 Hz; 360 x 295 x 175 mm
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

How the Titration Runs

In the Karl Fischer reaction, iodine reacts with water in the presence of sulphur dioxide, a base and an alcohol. The volume of reagent used is therefore a measure of the water. A dual platinum electrode sits in the titration vessel, which has a PTFE cover. While water remains, the iodine is used up at once. When free iodine appears, the current between the platinum pins rises sharply, the LED lights and the alarm sounds.

You key in the sample weight or density on the numeric keypad and start the run. The titrator reads the dispensed volume and works out the result itself. A port for an external serial printer lets you print any stored result later.

Practicals in Pharmaceutical Analysis

A standard practical is the water content of a drug powder or a solvent. Students weigh the sample on an Analytical Balance, add it to dried Methanol in the vessel and titrate. Before that, the class finds the factor of the reagent with a weighed drop of water.

A second exercise compares the Karl Fischer figure for one powder with its loss on drying from an Infra Red Moisture Balance. Drying also removes other volatile matter, so the two figures rarely agree. Dried samples wait in a Desiccator Plain until they are weighed.

Frequently Asked Questions

In which units does it report moisture?
It shows per cent, ppm or mg H2O. The titrator works the figure out from the reagent volume, the titer factor and the sample weight or density you keyed in.

How does it find the titer factor of the reagent?
Titrate a known mass of water, or a standard such as sodium tartrate dihydrate. The instrument calculates the factor itself and averages repeated runs, so one stray run has less effect on the value.

What is the end point wait time for?
Near the end, water held inside some samples is released slowly, so the end point can appear and then fade. The wait time, adjustable from 1 to 59 seconds, is how long it must hold before the titration counts as finished.

Which samples can it test?
Liquids, powders, fine particles and fibre samples that dissolve in alcohol or other solvents. The sample must release its water into the solvent in the vessel for the reagent to reach it.

Why do my results read high on humid days?
Moisture from room air adds to the water found. Keep the vessel closed, keep fresh calcium chloride in its bottle as the drying agent and do not leave the reagent bottles open.

Other Moisture Testing Instruments

For a quote, use Add To Cart or send us a message through the contact page.