Siemens 900C Ventilator

Product Code : EL-MV-3954
Jlab Export is a leading Siemens 900C Ventilator Manufacturer,and suppliers in India

The Siemens 900C is an intensive care ventilator fed by compressed gas. Gas reaches it from a central supply, a compressor or a cylinder through the high pressure inlet, or from flow meters through the low pressure inlet. JLab supplies the unit for hospitals and for biomedical engineering labs that study ventilator gas supply. It uses about 40 W of mains power.

Specifications

BrandSiemens
ConditionNew
High pressure gas inletCentral gas supply, compressor or gas cylinder
Electronic gas supply unit inlet2.5–5 bar (250–500 kPa, 35–70 psi)
Other gas supply inlet2.5–7 bar (250–700 kPa, 35–100 psi)
Low pressure gas inletVia flow meters
Gas consumptionAbout equal to the preset inspiratory minute volume or the flow meter setting
Power supply(100), 110, 120, 220 or 240 V, 50 or 60 Hz
Power consumptionApprox. 40 W
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
Supplied byJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Gas Supply Practicals for Biomedical Students

Students trace the gas path from the high pressure inlet to the patient circuit and name each part on the way. They connect an air source such as the Air Compressor Ventilators unit, measure the supply pressure and compare it with the inlet range in the table. A cylinder exercise uses a Cylinder Trolley to hold the cylinder upright beside the machine.

With an Anti-Static Rebreathing Bag on the patient end, the class runs a leak check and then lets the supply pressure fall to see how the machine responds.

Frequently Asked Questions

Which gas sources can feed it?
The high pressure inlet takes gas from a central gas supply, a compressor or a gas cylinder. The low pressure inlet takes gas through flow meters. Most hospitals run it from the wall pipeline and keep cylinders as a backup.

What inlet pressure does it need?
With the electronic gas supply unit, the inlet pressure is 2.5–5 bar (35–70 psi). Other gas supply inlets accept 2.5–7 bar (35–100 psi). A supply below these figures cannot drive the machine as intended.

What should be checked when the gas supply pressure drops?
Check the pipeline outlet or the cylinder contents first, then the hose and its connector. A clogged inlet filter or a full water trap on an air line also cuts the pressure. Faults inside the machine are a job for the biomedical team.

Which mains supplies does it accept?
It can be set for 110, 120, 220 or 240 V, with 100 V as an option, at 50 or 60 Hz. Power use is about 40 W, so it runs from an ordinary hospital socket.

How much gas does it use?
Gas use is roughly equal to the preset inspiratory minute volume, or to the flow meter setting when the low pressure inlet is used. That figure helps estimate how long a cylinder will last.