The Respironics Lifecare is a volume ventilator with five modes, from Control to SIMV, and tidal volumes from 0.05 to 3.00 litres. JLab supplies it with a battery test switch for the internal and external batteries, AC and DC circuit breakers, a low pressure alarm and a manual sigh button. Its switches and breakers suit power drills in a training lab.
Specifications
| Brand | Respironics |
| Condition | New |
| Modes | Control, Control + Sigh, Assist/Control, Assist/Control + Sigh, SIMV |
| Tidal volume | 0.05–3.00 litres |
| Sigh | 150% of set tidal volume, 1 sigh breath every 100 breaths; manual sigh button |
| Breath rate | 2–40 BPM |
| Peak inspiratory flow | 10–120 LPM |
| Airway pressure limit | 5–100 cm H2O |
| Sensitivity | +18 cm H2O to at least -6 cm H2O |
| Low pressure alarm | 2–50 cm H2O |
| Power controls | Battery test switch (internal/external), main power switch with recharge, AC and DC push/pull circuit breakers |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Supplied by | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Alarm and Power Practicals for Students
Students connect the circuit to an Anti-Static Rebreathing Bag as a test lung and run the machine in Control mode. Pulling the bag off the circuit sets off the low pressure alarm, and they note how long it takes. They then use the battery test switch on the internal and the external battery, and trip and reset the AC and DC breakers.
Home-care nursing classes add airway care to the drill. Students set up a Suction Tube and practise on the Multi-functional Airway Management Model, then reconnect the circuit and confirm that the alarm clears.
Frequently Asked Questions
How do I check the internal and external batteries?
Use the battery test switch, which tests the internal or the external battery on demand. Run the test at every shift check and before the unit moves, and keep it on mains with the switch at recharge when it is not in use.
What does the low pressure alarm tell me?
It sounds when the pressure in the circuit stays below the alarm limit, set from 2 to 50 cm H2O. The usual causes are a disconnected or leaking circuit. Check the patient end and every joint first.
What does the manual sigh button do?
Pressing it gives one sigh breath. In the sigh modes, the machine also gives a sigh of 150% of the set tidal volume once every 100 breaths. Trained staff decide whether sighs are used.
What should I do if a circuit breaker trips?
The AC and DC breakers are push/pull types. Find the cause first, such as a faulty lead or battery, then push the breaker back in. A breaker that trips again means the unit needs a biomedical technician.
Who sets the mode and tidal volume?
Modes, volumes and rates are set by trained clinical staff under the hospital protocol or the prescriber, never by a caregiver on their own. Caregivers and students learn to read the settings and respond to alarms.