Drager Evita Ventilators Manufacturer,Supplier and Exporter in India
Product Code : EL-MV-3968
Jlab Export is a leading Drager Evita Ventilators Manufacturer,and suppliers in India
On its 6.5 inch colour screen, the Drager Evita shows peak, plateau, mean and minimum airway pressure, PEEP, minute volume and tidal volume. Its alarms watch airway pressure, expired minute volume, tidal volume, apnea and spontaneous breath rate. It runs on 100–240 V mains, with an optional DC supply, and JLab supplies it to hospitals and training departments.
Specifications
| Brand | Drager |
| Condition | New |
| Monitored airway pressure | Peak, plateau, mean, PEEP and minimum, 0 to 99 mbar (cmH2O) |
| Monitored minute volume (BTPS) | Total, spontaneous and leak, 0 to 99 L/min |
| Monitored tidal volume (BTPS) | Inspired, expired and pressure-support, 0 to 3999 mL |
| Alarms | Airway pressure high/low, expired minute volume high/low, tidal volume high, apnea time 5 to 60 s, spontaneous breath frequency high |
| Maximum flow, pressure support and spontaneous breathing | 180 L/min adult, 60 L/min paediatric |
| Power | 100–240 V, 50/60 Hz, approx. 125 W, optional 10–30 V DC |
| Gas supply pressure, O2 and air | 2.7–6 bar (39–87 psi) |
| Size and weight | 530 x 290 x 450 mm (W x H x D, 20.9 x 11.4 x 17.7 in) without trolley, basic unit approx. 27 kg (60 lb) |
| Display | 6.5 in colour TFT LCD |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Supplied by | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Alarm Testing Practicals for Students
Biomedical-engineering students can read every measured value off the screen while the ventilator breathes into an Anti-Static Rebreathing Bag used as a test lung. Squeezing the bag or pinching the tubing shows which pressure and volume alarms respond. For the apnea alarm, they time the delay after the last breath with a Digital Stopwatch and compare it with the time set on the panel.
Where the DC option is fitted, students check the external supply with a Digital Multimeter before connecting it. Nursing students can practise reading the gap between inspired and expired tidal volume, which points to a leak in the circuit.
Frequently Asked Questions
What is the difference between peak and plateau pressure on the screen?
Peak pressure is the highest airway pressure reached during the breath. Plateau pressure is read during a short pause at the end of inspiration, when no gas is flowing. The Evita shows both, along with mean pressure, PEEP and minimum pressure, on a 0 to 99 mbar scale.
What does the apnea alarm time control?
It sets how long the ventilator waits without a breath before it raises the apnea alarm. On this unit the time can be set from 5 to 60 seconds. The value is chosen by trained staff under the hospital protocol.
Why does the screen show both inspired and expired tidal volume?
Comparing the gas that goes in with the gas that comes back shows whether some is escaping. A gap between the two figures points to a leak at a cuff, mask or circuit joint. The Evita also shows a separate leak minute volume.
What does the high spontaneous breath frequency alarm watch?
It counts the breaths the patient takes on their own and alarms when that rate rises above the limit staff have set. It draws attention to fast breathing that the set rate alone would not show.