Specific Heat of Air Apparatus

Product Code : EL-HTL-10047
Jlab Export is a leading Specific Heat of Air Apparatus Manufacturer,and suppliers in India

Specific Heat of Air Apparatus (code EL-HTL-10047) is a flow rig in which a blower drives air over a 1 kW heating coil placed centrally in a test section, for finding the specific heat of air from the heater input, the air flow and the temperature rise. An orifice meter measures the flow and a multichannel digital indicator, reading 0–300 °C, shows the air temperatures. JLab builds the blower, test section and meter panel of this specific heat apparatus in Ambala, India.

Specifications

ParameterDetail
BlowerAir blower driven by an FHP motor
Air flow measurementOrifice meter
Heater1 kW heating coil placed centrally in the test section
Temperature measurementMultichannel digital indicator, 0–300 °C
Input measurementVoltmeter and ammeter
Electricity220 V single phase stabilised AC
Floor space2 m x 1 m x 2 m height
Quality checkEach specific heat apparatus is visually inspected before dispatch
Pack quantityEach (sold individually)
Minimum order quantity1 unit
Dispatch timeUsually 14–18 working days
Product codeEL-HTL-10047
ManufacturerJain Laboratory Instruments Pvt Ltd (JLab) — OEM
Country of originIndia (made in Ambala, Haryana)

The Steady-Flow Energy Balance

Once the outlet temperature is steady, the electrical input goes into heating the moving air: VI = ṁ cp (T2 − T1), so cp = VI/[ṁ(T2 − T1)]. The orifice meter gives the volume flow of air, and multiplying it by the air density at the inlet temperature and room pressure gives the mass flow ṁ. Because the air passes through at almost atmospheric pressure, the value found is the specific heat at constant pressure, cp, the figure used in steady-flow work on ducts and heat exchangers.

Getting a Reliable Value

Some heat escapes through the walls of the test section instead of reaching the air, which makes the calculated cp come out high. Runs at several air flows and heater settings show how large that loss is: at higher flow the temperature rise is smaller, the walls stay cooler and the loss becomes a smaller share of the input. Taking the inlet and outlet readings only after they settle avoids counting heat that is still warming the duct.

Other Rigs With Metered Air Flow

Browse more Heat Transfer Lab rigs under Engineering Lab Equipment.