Stefan Boltzmann Apparatus

Product Code : EL-HTL-10049
Jlab Export is a leading Stefan Boltzmann Apparatus Manufacturer,and suppliers in India

Stefan Boltzmann Apparatus (code EL-HTL-10049) is a radiation set-up for determining the Stefan-Boltzmann constant: a blackened 20 mm copper disc is inserted at the centre of a 200 mm copper hemisphere kept hot by a 250 mm water jacket, and the rate at which the disc warms gives σ. A buzzer timer sounds every 5 seconds for the disc readings. The copper hemisphere, water jacket and test disc are made by JLab in Ambala, India.

Specifications

ParameterDetail
HemisphereCopper sheet, 200 mm diameter
Water jacket250 mm diameter, surrounding the hemisphere
Hot water supplyWater heating tank with an electric immersion heater
Test discCopper, 20 mm diameter, blackened, with a thermocouple at its centre
Temperature indicatorMultichannel digital, 0–200 °C, least count 0.1 °C, for hemisphere and disc
TimerAudible buzzer sounding every 5 seconds
Supplied withTechnical manual
Electricity230 V, 15 A AC, with earthing
Bench areaAbout 1 m x 0.5 m at working height
Quality checkEach radiation apparatus is visually inspected before dispatch
Pack quantityEach (sold individually)
Minimum order quantity1 unit
Dispatch timeUsually 14–18 working days
Product codeEL-HTL-10049
ManufacturerJain Laboratory Instruments Pvt Ltd (JLab) — OEM
Country of originIndia (made in Ambala, Haryana)

From Disc Warming Rate to σ

Hot water from the immersion-heated tank fills the jacket until the hemisphere settles at a steady temperature Th. The disc, still at room temperature Td, is then inserted and its centre thermocouple is read at each buzzer signal. Over the first few readings, radiation from the hemisphere is practically the only heat reaching the disc, so the energy balance is m c (dTd/dt) = σ A (Th⁴ − Td⁴), with m and c the mass and specific heat of the copper disc and A = πd²/4 its exposed face. The slope of the temperature–time plot at the moment of insertion then gives σ = m c (dTd/dt)/[A(Th⁴ − Td⁴)], all temperatures in kelvin.

Why the Timer and 0.1 °C Resolution Matter

The useful part of the record is short, because the balance holds most closely while the disc is still near room temperature. The buzzer keeps the readings evenly spaced at 5 seconds, and the 0.1 °C least count resolves the small rise between them, so the initial slope can be drawn from a handful of points. Results are judged against the textbook figure for σ of about 5.67 × 10⁻⁸ W/m²K⁴.

Other Radiation Set-ups

More radiation and conduction rigs are in Heat Transfer Lab, part of Engineering Lab Equipment.

Request a Quote

Quote requests go through the contact page or Add To Cart; include the number of sets and the town or port of delivery.

Last updated: 30 October 2026


   

Office Address

Works: 2475-84,
Hargolal Road,Ambala, Haryana

[email protected]

+91-85699-09696

Follow Us