Lagged Pipe Apparatus (code EL-HTL-10042) is a conduction rig of three concentric G.I. pipes, 1 m long, with one lagging material filling the gap between the inner and middle pipes and a second lagging in the outer gap. A central cartridge heater drives heat outwards, and thermocouples on the pipe surfaces give the temperatures for finding the conductivity of each lagging. JLab builds the pipe assembly, stand and meter panel in Ambala, India.

Specifications

ParameterDetail
PipesThree concentric G.I. pipes, 1 m long, on a stand
Innermost pipe6 cm outer diameter
Middle pipe8.5 cm mean diameter
Outermost pipe10.7 cm inner diameter
LaggingOne material between the inner and middle pipes, a second between the middle and outer pipes
HeaterNichrome wire cartridge heater placed centrally along the pipe
Control panelSingle phase dimmerstat; voltmeter 0–250 V; ammeter 0–2 A
Temperature measurementMultichannel digital indicator, 0–300 °C, with Cr-Al thermocouples on the pipe surfaces
ElectricitySingle phase 230 V AC
Floor area1.5 m x 1 m x 0.5 m high
Quality checkEach lagged pipe apparatus is visually inspected before dispatch
Pack quantityEach (sold individually)
Minimum order quantity1 unit
Dispatch timeUsually 14–18 working days
Product codeEL-HTL-10042
ManufacturerJain Laboratory Instruments Pvt Ltd (JLab) — OEM
Country of originIndia (made in Ambala, Haryana)

Two Laggings in Series

At steady state the heater output, Q = VI, crosses both laggings in turn. A cylindrical layer between radii ra and rb over a length L has a resistance of ln(rb/ra)/(2πkL), so the inner lagging gives k1 = Q ln(r2/r1)/[2πL(T1 − T2)] and the outer one k2 = Q ln(r3/r2)/[2πL(T2 − T3)]. Here r1 comes from the 6 cm innermost pipe, r2 from the 8.5 cm middle pipe and r3 from the 10.7 cm bore of the outer pipe. The thin G.I. walls add almost no resistance, which is why the middle pipe is given by its mean diameter: its one surface temperature, T2, serves both laggings.

What the Temperature Profile Shows

Each layer takes a share of the overall fall from T1 to T3 in proportion to its resistance, ln(rb/ra)/(2πkL), so the drop across a layer grows as its conductivity falls or its radius ratio rises. Runs at two or three dimmerstat settings show whether k changes with mean temperature, as it often does for insulating materials, and the same readings give a combined conductivity for both laggings treated as one layer between r1 and r3.

Other Radial and Flat-Wall Conduction Rigs

Find it with the other Heat Transfer Lab rigs in Engineering Lab Equipment.

Request a Quote

Send the quantity and the delivery town through the contact page, or add the apparatus to an enquiry with Add To Cart, and JLab will reply with a quotation.

Last updated: 30 October 2026


   

Office Address

Works: 2475-84,
Hargolal Road,Ambala, Haryana

[email protected]

+91-85699-09696

Follow Us