Thermal Conductivity of Insulating Slabs (code EL-HTL-10054) is a guarded hot plate apparatus for measuring how readily an insulating material conducts heat. Two slab specimens of equal thickness sit on either side of a 110 mm main heater ringed by a radial guard heater, with water-cooled plates on their outer faces, all inside an insulated enclosure. JLab manufactures this two-slab conductivity apparatus in Ambala, India.

Specifications

ParameterDetail
Main heaterPlate 110 mm dia., mica heater sandwiched between copper plates
Radial guard heaterPlate ID 120 mm, OD 200 mm, mica heater between copper plates
Cooling platesWater circulated, 2 nos.
Heater controlDimmerstat, 2 A, 2 nos., one for each heater
InstrumentsVoltmeter and ammeter with selector switches; multichannel digital temperature indicator, 0.1 °C least count
Services230 V AC stabilised supply with earthing; water about 4–5 litre/min; bench area 1 m x 1 m x 0.5 m height
DocumentationTechnical manual
Quality checkEach slab apparatus is visually inspected before dispatch
Pack quantityEach (sold individually)
Minimum order quantity1 unit
Dispatch timeUsually 14–18 working days
Product codeEL-HTL-10054
ManufacturerJain Laboratory Instruments Pvt Ltd (JLab) — OEM
Country of originIndia (made in Ambala, Haryana)

Why the Guard Heater Matters

The stack is symmetrical: cooling plate, specimen, heater plates, specimen, cooling plate. If the edge of the main heater plate were hotter than its surroundings, part of its heat would leak sideways and the result would read too high. The radial guard heater, run on its own dimmerstat, is brought to the same temperature as the main heater plate, so there is no temperature difference across the gap between them. All the main heater input then flows axially through the two slabs to the cooling plates, giving the one-dimensional heat flow that Fourier’s law assumes.

Calculating k From the Readings

The selector switches put the voltmeter and ammeter on the main heater, so Q = V × I. Half of Q passes through each slab, so k = Q t / (2 A (Th − Tc)), where t is the slab thickness, A the face area of the 110 mm main heater plate, and Th and Tc the hot and cold face temperatures. Insulating slabs take a long time to settle, so readings are logged until the temperatures stop changing; the 0.1 °C least count shows when they have.

Other Thermal Conductivity Apparatus

All four sit with this apparatus in Heat Transfer Lab, a section of Engineering Lab Equipment.

Request a Quote

To order, add the apparatus through Add To Cart or give the quantity and the delivery address on the contact page, and a quotation will follow.

Last updated: 30 October 2026


   

Office Address

Works: 2475-84,
Hargolal Road,Ambala, Haryana

[email protected]

+91-85699-09696

Follow Us