UV Spectrophotometer Manufacturer,Supplier and Exporter in India
Product Code : EL-S-2166
Jlab Export is a leading UV Spectrophotometer Manufacturer,and suppliers in India
The large sample compartment sets this UV spectrophotometer apart. With optional holders it takes cuvettes from 5 to 100 mm path length, so very dilute samples can be read in a long cell. It covers 200 to 1000 nm with a 5 nm bandwidth and sets 0 and 100 %T by itself. The standard curve appears on a 128 x 64 dot graphical LCD.
Specifications
| Monochromator | CT grating, 1200 lines/mm |
| Wavelength range | 200–1000 nm |
| Spectral bandwidth | 5 nm |
| Wavelength accuracy | ±2 nm, repeatability 1 nm |
| Photometric accuracy | ±0.5 %T, repeatability 0.2 %T |
| Display range | 0–200 %T, -0.3 to 3 A |
| Measurement modes | Transmittance (T), absorbance (A) and concentration (C) |
| Zero setting | Automatic 0 and 100 %T |
| Display | 128 x 64 dot graphical LCD with on-screen standard curve |
| Sample compartment | 4-position 10 mm cell changer supplied; 5–100 mm cuvettes with optional holders |
| Software and outputs | Basic PC software (quantitative, kinetic) supplied; USB; Centronics printer port |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Water Analysis and Kinetics Practicals
Environmental chemistry students can estimate nitrate in drinking water from its absorbance in the UV. Filtered samples are read in quartz cuvettes against a set of nitrate standards made up in Volumetric Flasks, Class A Borosilicate Glass Plastic Stopper JLab. Each aliquot is drawn with Bulb Pipettes, Class B Soda Glass, One Mark JLab fitted with a Pipette Filler, Bulb, never by mouth.
The Basic PC software runs a kinetic mode. A teacher can follow an enzyme reaction or the fading of a dye on a computer screen in front of the class. The quantitative mode keeps the standard curve, and the samples are read against it in the same session.
Frequently Asked Questions
When should I use a long-path cuvette?
Use one when a sample absorbs too weakly to read well in a 10 mm cell. By the Beer-Lambert law, absorbance rises in step with path length. A 100 mm cell gives ten times the absorbance of a 10 mm cell for the same solution.
What does the automatic zero setting do?
With the blank in place, the instrument sets 0 %T and 100 %T by itself, so nobody turns a knob between readings. Run it again whenever you change the wavelength, the solvent or the reagent blank.
Why does a sample sometimes read above 100 %T?
It means the sample passes more light than the blank did. The blank may hold a reagent the sample lacks, or the cuvettes may not be matched or clean. The display runs to 200 %T, so such a reading is shown rather than cut off.
Which PC software comes with it?
The Basic software for quantitative and kinetic work is included. The Professional software is optional and adds wavelength scanning and DNA/protein analysis to the same instrument.
Is a 5 nm bandwidth narrow enough?
For routine quantitative work it is, because most solutions in teaching labs have broad absorption bands. Narrow, sharp peaks are better measured on a 2 nm bandwidth model, which follows their shape more closely.
Other Models Taking Long-Path Cells
- Microprocessor Spectrophotometer (visible, 5–100 mm cuvettes)
- Visible Spectrophotometer (100 mm optical path)
- Digital Spectrophotometer (provision for longer test tubes)
For a quote, use Add To Cart or send us a message through the contact page.