Surgical Cautery Unit

Product Code : EL-SE-3855
Jlab Export is a leading Surgical Cautery Unit Manufacturer,and suppliers in India

An electrosurgical or cautery unit passes high-frequency current through an active electrode to cut tissue or to coagulate it. This one has monopolar cut, blend and coagulation outputs and a separate bipolar output for cutting and coagulation. Either a hand switch or the PVC foot pedal fires it, and each mode has its own audio-visual alarm.

Specifications

Monopolar modesCut, coagulation and blend
Bipolar modesCut and coagulation
ActivationHand switch or foot switch, independently
AlarmAudio-visual, separate for each mode
Foot pedalPVC
Monopolar generator400 ±25 kHz; rated cut 250 W, blend 175 W, coagulation 160 W into 500 Ω
Bipolar generator650 ±25 kHz; rated coagulation 80 W into 125 Ω
DimensionsW 285 × D 295 × H 120 mm
Power supply220 V AC, 50 Hz; 750 VA consumption
Gross weightAbout 20 lb
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Biomedical Lab Exercises on the Unit

Trainees first name every control and socket on the front panel and draw the current path for each output. In monopolar work, current runs from the active electrode through the patient to the return plate. In bipolar work, it passes only between the two tips of the forceps.

With the unit unplugged, students check the foot pedal lead and the hand switch for continuity with a Digital Multimeter, pressing each switch in turn. Nursing students practise placing a patient return plate on a fleshy area of a Multi Functional Patient Care Nursing Manikin and routing cables clear of the floor. The session ends with wiping the pedal and coiling the leads.

Frequently Asked Questions

What is the difference between the monopolar and bipolar outputs?
Monopolar current flows from a small active electrode through the body to a large return plate. Bipolar current flows only between the two tips of bipolar forceps, so the effect stays between the tips and no return plate is used.

What do the cut, blend and coagulation controls change?
Cut gives a continuous waveform that parts tissue along the electrode. Coagulation gives short bursts that heat and dry tissue to stop small bleeders. Blend mixes the two. Output levels for clinical use follow the hospital’s own protocol and trained staff.

Why does monopolar work need a large patient plate?
The plate spreads the returning current over a wide area of skin, so the current density there stays low and the skin does not heat. If part of the plate lifts, the current crowds into a small area, which is how plate-site burns happen.

What usually goes wrong with the foot pedal?
Most faults are a damaged lead where it enters the pedal, dirty or bent connector pins, and fluid or dirt inside the pedal. Inspect and clean it after each list, and check continuity with the unit switched off and unplugged.

Why does each mode have its own sound and light?
The alarm tells everyone which output is live while the switch is pressed. If a tone or lamp fails, take the unit out of use and send it to the biomedical department for checking.