Travelling Microscope (code EL-PL-12377) is a microscope on two sliding carriages, one moving vertically and one horizontally along stainless steel scales, each read by a vernier with a least count of 0.01 mm. It measures small distances, such as the bore of a capillary or the height of a meniscus, by sighting each end in turn. JLab manufactures the instrument in Ambala, India, and packs it in a polished wooden case.
Specifications
| Parameter | Detail |
|---|---|
| Focusing | Rack and pinion on the microscope tube |
| Eyepiece | Ramsden, with cross-line graticule |
| Carriages | Vertical carriage on a brass pillar; transverse carriage on a metal rod in the base |
| Scales | Horizontal and vertical, stainless steel |
| Vernier least count | 0.01 mm on each carriage |
| Adjustment | Locking screws for coarse setting; fine adjustment knob on each slide |
| Vernier reading | Provision for a swing-out magnifier |
| Base | Heavy cast metal with levelling screws; removable flat horizontal stage |
| Packing | Polished wooden case |
| Quality check | Each travelling microscope is visually inspected before dispatch |
| Pack quantity | Each (sold individually) |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Product code | EL-PL-12377 |
| Manufacturer | Jain Laboratory Instruments Pvt Ltd (JLab) — OEM |
| Country of origin | India (made in Ambala, Haryana) |
How the Carriages Move
The microscope tube, focused by rack and pinion, rides on a vertical carriage that runs up a brass pillar. The pillar stands on a transverse carriage that slides along a metal rod in the base beside the horizontal scale. For a coarse setting, slacken the locking screw and push the carriage near the mark; tighten it, then bring the cross-line onto the mark with the fine adjustment knob at the end of that slide. The microscope tube itself can be adjusted vertically as well as horizontally.
Taking a Reading
Level the base with its levelling screws and focus the Ramsden eyepiece on the cross-lines before focusing on the object, so the lines show no parallax against the image. Set the cross-line on one edge and add the main scale reading to the coinciding vernier division times 0.01 mm. Move to the far edge, turning the fine adjustment knob in one direction only so that backlash does not creep in, and subtract the two readings. A swing-out magnifier can be attached to read the vernier.
Practicals It Serves
In the capillary rise method it reads the meniscus height and the bore of the tube. With the tube looking down at the removable stage, it finds the refractive index of a glass slab: focus on a mark, then on the mark seen through the slab, then on the top face, and divide the real depth by the apparent depth.
Related Measuring Apparatus
- Travelling Microscope — another model, with two slow-motion knobs and a milky stage sheet.
- Rising Table — raises and lowers the liquid for capillary rise readings.
- Capillary Tube — borosilicate capillaries of different bores to measure.
Other measuring and optics apparatus is in the Physics Lab range.