Automatic Distillation Unit Manufacturer,Supplier and Exporter in India
Product Code : EL-MLE-13269
Jlab Export is a leading Automatic Distillation Unit Manufacturer,and suppliers in India
Automatic Distillation Unit (code EL-MLE-13269) is an automatic analyser for the ASTM D86 distillation of petroleum products that controls the heating, tracks the recovered volume with a laser and plots the distillation curve on a 10-inch touch screen. It reports the initial and final boiling points and the temperature at each percentage recovered. The analyser is built in Ambala, India, by JLab.
Specifications
| Parameter | Detail |
|---|---|
| Test method | ASTM D86 |
| Display | 10-inch touch LCD: temperature, volume and curves |
| Volume tracking | Laser level tracker on stepping linear motor and ball screw |
| Cooling | Mechanically refrigerated cooling tube and receiver chamber; 0.5 kW, 0–100 °C |
| Distillation rate | 4–5 ml per minute from initial boiling point to 95% recovered |
| Heating power | 2 kW |
| Vapour temperature range | 0–400 °C |
| Furnace temperature range | 0–500 °C |
| Volume accuracy | ±0.1 ml |
| Results | Stored, queried and printed; corrected to standard atmospheric pressure |
| Quality check | Each distillation unit is visually inspected before dispatch |
| Pack quantity | Each (sold individually) |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Product code | EL-MLE-13269 |
| Manufacturer | Jain Laboratory Instruments Pvt Ltd (JLab) — OEM |
| Country of origin | India (made in Ambala, Haryana) |
What the Distillation Curve Records
In this atmospheric distillation the sample is boiled by the electric furnace, its vapour is condensed in the refrigerated cooling tube, and the condensate collects in the receiver. Vapour temperature is logged against the volume collected: the first drop of condensate gives the initial boiling point, the highest temperature reached gives the final boiling point, and the points between them form the curve that describes the volatility of the fuel.
Automatic Control and Pressure Correction
The heater is regulated so that, from the initial boiling point to 95% recovered, the distillation rate stays within 4–5 ml per minute. The unit measures local atmospheric pressure and corrects the readings to standard atmospheric pressure, and each result can be stored, recalled and printed.
Other Fuel and Oil Test Apparatus
- Copper Strip Corrosion Tester — heated metal bath taking 4 to 12 copper strip corrosion samples.
- Conradson — carbon residue tester with porcelain and iron crucibles and chimney.
- Flash Point — Cleveland open-cup flash and fire point apparatus with electric heating.
- Cloud and Pour Point — cold-property tester with two individually controlled refrigerated compartments.
The analyser belongs to Mechatronics Lab Equipments, a section of Mechanical Engineering Lab Equipment.
Request a Quote
Use Add To Cart to request a quotation, or write through our contact page with the quantity, your mains voltage and the delivery town.
Last updated: 30 October 2026