Hand Held Single Channel ECG

Product Code : EL-EM-3940
Jlab Export is a leading Hand Held Single Channel ECG Manufacturer,and suppliers in India

A single-channel ECG records the heart’s electrical activity one lead at a time, and this unit is small enough to hold during the recording. Leads are picked automatically or by hand. It has a floating input, a 1 mV calibration pulse, a 0.05–150 Hz response and sensitivity steps of 5, 10 and 20 mm/mV.

Specifications

Input circuitFloating
Calibration voltage1 mV
CMRRAbove 103 dB
A/D conversion12 bit
Frequency response0.05–150 Hz
FiltersAC 50/60 Hz, EMG 35–45 Hz
Skin voltage tolerance300 mV
Sensitivity5, 10 and 20 mm/mV
Noise level15 µVp-p or less
Lead selectionAutomatic and manual
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Practicals for Nursing and Biomedical Students

Nursing students place the four limb electrodes and the chest electrodes on a Multi-Functional Nursing Manikin and name each lead as the unit steps through them. On a seated classmate, they record lead II at 10 mm/mV, then at 5 and 20 mm/mV, and measure how the calibration pulse changes height.

Biomedical engineering students test each lead wire end to end with a Digital Multimeter to find broken cores. They then record near a mains cable with the AC filter off and on, and compare the baseline. The session ends with cleaning the electrodes and coiling the patient cable loosely.

Frequently Asked Questions

What does the 1 mV calibration square show?
It shows how tall 1 mV will appear on the trace. At 10 mm/mV the square should be 10 mm high, at 5 mm/mV half that and at 20 mm/mV double. A square of the wrong height means the sensitivity setting needs checking before the trace is read.

When should the EMG filter be switched on?
Use it when muscle tremor from a cold, tense or shivering subject puts fine fuzz on the trace. The 35–45 Hz filter removes much of that tremor, but it also softens sharp parts of the waveform, so note on the record that it was on.

Why are the lines thick and blurred even with the filters on?
The usual cause is poor skin contact. Clean oil and lotion from the skin, use fresh electrodes that stick well, and keep the patient cable away from mains leads. Ask the subject to relax and breathe quietly during the recording.

What is the difference between automatic and manual lead selection?
In automatic mode the unit steps through the leads by itself. In manual mode the user picks each lead, which suits a practical where one lead is studied at a time or a single lead must be repeated.