Low Temperature Cabinets Manufacturer,Supplier and Exporter in India
Product Code : EL-LTC-2320
Jlab Export is a leading Low Temperature Cabinets Manufacturer,and suppliers in India
These low temperature cabinets freeze and hold blood components, serum, vaccines and clinical specimens. Each one is double walled, with powder-coated sheet steel outside, a stainless steel interior and PUF insulation between. A digital indicator-cum-controller sets the temperature, and an inner acrylic door holds the cold in while the counter balanced outer door is open.
Specifications
| Use | Quick freezing and storage of blood components, serum, vaccines and specimens |
| Temperature versions | −20 °C, −30 °C, −70 °C |
| Designs | Horizontal or vertical |
| Capacities | 55, 120, 170, 280, 425 and 650 L (2, 4.5, 6, 10, 15 and 23.5 cu ft) |
| Body | Double walled, powder-coated sheet steel outside, stainless steel inside, PUF insulation |
| Refrigeration | Capillary controlled, balanced, hermetically sealed |
| Control | Digital electronic indicator-cum-controller |
| Doors | Counter balanced outer door, inner acrylic door |
| Power | 220 V AC, 50 Hz, single phase |
| Option | Digital indicator-cum-controller with computer data logger |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Training Exercises for Lab and Biomedical Students
Biomedical-engineering students begin by tracing the parts: the hermetic compressor, the capillary line, the controller and its sensor, and the door gaskets. A simple seal check follows. They close the outer door on a strip of paper at several points around the gasket and feel how firmly it is held.
For temperature practice on a −20 °C or −30 °C cabinet, students read the centre of the chamber with a Digital Stem Thermometer and compare it with the controller display. They log both figures at set times through the day. A Temperature and Humidity Data Logger placed beside the cabinet records the room conditions, so they can see how a warm afternoon affects the run time.
Frequently Asked Questions
Why is the outer door hard to open again just after it closes?
Warm air that enters while the door is open cools and shrinks once it is shut, which leaves a slight vacuum inside. Wait a short while for the pressure to even out instead of forcing the handle.
How do I deal with frost on the door gasket?
Wipe it off with a soft dry cloth or a plastic scraper, never a knife or metal tool. Frost on the gasket stops the door sealing, which lets in more moist air and builds more frost.
Does the cabinet need its own power socket?
Yes. Run it from a dedicated, earthed 220 V AC, 50 Hz single phase socket, not from a multi-plug shared with heaters or other machines. A loose or overloaded supply can stop the compressor without warning.
How do I check that the displayed temperature is right?
Place a calibrated low-temperature thermometer probe in the centre of the chamber, close the door and read it once the figure is steady. Compare it with the display and record both. A gap larger than the lab allows calls for a service check.
Other Cold Storage Units
For a quote, use Add To Cart or send us a message through the contact page.
Last updated: 13 November 2026