• +91-85699-09696

Blood Bag Sealer Manufacturer,Supplier and Exporter in India

Blood Bag Sealer

Product Code : EL-BBS-2313
Jlab Export is a leading Blood Bag Sealer Manufacturer,and suppliers in India

This bench unit closes the PVC tube of a blood bag with a seal made between two electrodes. The cycle begins by itself once the tube lies between the high grade metal electrodes, and four LEDs report power, ready, seal and ground. The caps in front of the electrodes lift off, so the sealing area can be wiped clean.

Specifications

StartAutomatic, once the tube lies between the electrodes
ElectrodesHigh grade metal
Electrode capsRemovable for cleaning
LED indicatorsPower, Ready, Seal, Ground
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Workshop Practice in the Blood Bank Lab

Biomedical-engineering students start with a function check. They power on, wait for the Ready LED, then seal a piece of empty tubing and pull on it to test the joint. They take the electrode caps off, clean them, refit them and make a second test seal to confirm the head was put back right.

In a mock collection station set up beside a Donor Couch, MLT students seal the tube of a water-filled practice bag close to the bag. They can also seal matching tubing on a Blood Bag Tube Sealer Automatic and compare how the two seals look and pull.

Frequently Asked Questions

Can damp or wet tubing be sealed?
Dry the tube first. Moisture on the tubing or on the electrodes spoils the heating, so the seal may not form or may be weak. Wipe both dry, then seal again.

Why does the tube sometimes split during sealing?
Splitting usually happens when the tube is stretched or pulled while the plastic is soft, or when two seals are made too close together. Lay the tube in loosely and hold it still until the Seal LED goes out.

What should I check after refitting the electrode caps?
Make a test seal on spare empty tubing before sealing a bag. Look for a flat, even band across the tube and pull gently on both sides. A clean, firm seal shows the caps sit properly.

Where should the tube lie between the electrodes?
Lay it straight across the middle of the electrodes, a short way from the bag, with no twist or fold. The seal starts by itself once the tube is in place, so keep fingers clear of the electrodes.

Other Blood Collection Equipment

For a quote, use Add To Cart or send us a message through the contact page.

Last updated: 13 November 2026


   

Office Address

Works: 2475-84,
Hargolal Road,Ambala, Haryana

[email protected]

+91-85699-09696

Follow Us