Environmental Chamber Manufacturer,Supplier and Exporter in India
Product Code : EL-EGC-3895
Jlab Export is a leading Environmental Chamber Manufacturer,and suppliers in India
This cooled stability chamber keeps samples at a set temperature and relative humidity for quality control and storage tests. A refrigeration system lets it hold from 10 to 60 °C, below room temperature as well as above it. A steam generator raises the humidity anywhere from 40 to 95% RH, and a fan keeps both even through the stainless steel interior.
Specifications
| Type | Cooled stability chamber with refrigeration |
| Temperature range | 10–60 °C ± 1 °C |
| Humidity range | 40–95% RH ± 3% RH (from 5% above ambient) |
| Humidity source | Steam generation tank at base or top |
| Air circulation | Self-cooled fan |
| Inner chamber | Stainless steel, SS 304 grade |
| Outer body | Mild steel sheet, painted, on castor wheels |
| Insulation | Polyurethane foam (PUF) |
| Door | Double walled, magnetic gum packing, lock and key |
| Shelves | Adjustable |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Stability and Growth Practicals
In a pharmaceutics lab, students run a short accelerated stability study. They store tablets in open and closed packs at a raised temperature and humidity, and take out a few at each interval. Each pull is checked for appearance and weight, for crushing strength on the Tablet Hardness Tester and for break-up time on the Disintegration Test Apparatus. Comparing the packs shows how much moisture protection the packaging gives.
Because the chamber can run cooler than the room, botany and zoology classes also use it for seed germination or insect rearing at fixed conditions. A Temperature and Humidity Data Logger placed on a shelf records the conditions the samples actually saw during the study.
Frequently Asked Questions
Why does a stability chamber need cooling?
The fan and the steam generator both add heat, and the room itself may be warmer than the set point. The refrigeration system removes that heat, so the chamber can hold settings near or below room temperature, down to 10 °C.
Where does the humidity come from?
Water in a steam generation tank at the base or top is heated, and the steam is carried into the chamber by the circulation fan. Fill the tank with distilled water, since tap water leaves scale that slows steam output.
How should I load samples so conditions stay even?
Leave gaps between packs and between the load and the walls, so the fan can move air past every sample. Do not cover whole shelves with trays or boxes. The shelves adjust, so set their spacing to suit the pack sizes.
How do I keep the door seal working?
Wipe the magnetic gum packing clean and check that it closes flat all round. A worn or dirty seal lets moist air escape, and the humidity then drops each time the room air leaks in.
Other Climate and Growth Chambers
- Environmental Stability Chamber (inner acrylic door, stainless trays)
- Stability Chamber Microprocessor Control
- Carbon Di-Oxide Incubator
For a quote, use Add To Cart or send us a message through the contact page.