pH Tester, Checker Plus Type Manufacturer,Supplier and Exporter in India
Product Code : EL-PL-13545
Jlab Export is a leading pH Tester, Checker Plus Type Manufacturer,and suppliers in India
pH Tester, Checker Plus Type (code EL-PL-13545) is a compact hand-held tester, 71 x 50 x 21 mm, that reads 0 to 14 pH with 0.01 pH resolution and offers automatic two-point calibration. It runs on one CR 2032 button cell, and its HI 1271 general-purpose electrode mounts directly on the body. JLab supplies the Checker Plus tester from Ambala, India.
Specifications
| Parameter | Detail |
|---|---|
| Brand | Checker Plus |
| Measuring range | 0 to 14 pH |
| Resolution | 0.01 pH |
| Accuracy | ± 0.2 pH |
| Calibration | Automatic, 2-point |
| Dimensions | 71 x 50 x 21 mm |
| Power supply | 1 x CR 2032 button cell |
| Supplied with | HI 1271 pH electrode, battery and manual |
| Electrode HI 1271 | General use; plastic body Ø8 x 113 mm; viscolene electrolyte; 0 to 13 pH; 0 to 50 °C; mounts directly on the tester |
| Accessories | Preservative solution for pH and RedOx electrodes 500 ml; buffer pH 4.01, 7.01 and 10.01 (25 x 20 ml sachets each); spare CR 2032 battery; spare HI 1271 electrode |
| Quality check | Each pH tester is visually inspected before dispatch |
| Pack quantity | Each (sold individually) |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Product code | EL-PL-13545 |
| Supplied by | Jain Laboratory Instruments Pvt Ltd (JLab), Ambala, India |
Calibrating With the Buffer Sachets
For a two-point calibration, start with the pH 7.01 buffer and then use pH 4.01 for acidic samples or pH 10.01 for alkaline ones. Each 20 ml sachet is opened for one calibration, so the buffer is fresh each time and there is no stock bottle for a dirty electrode to contaminate. Rinse the electrode with distilled water between buffers and before each sample.
Electrode Range and Care
The display covers 0 to 14 pH, but the HI 1271 electrode is rated for 0 to 13 pH and 0 to 50 °C, so readings beyond those limits are outside what the electrode is meant for. Between uses, keep the electrode tip wet with the preservative solution rather than storing it dry or in distilled water. When a worn electrode no longer calibrates, the spare HI 1271 fits straight onto the tester.
Related pH Testing Items
- Electronic pH Tester — pocket-size tester with its own screw-type combination electrode.
- pH/Conductivity/TDS/Salt/Temp Waterproof Tester — waterproof tester that also reads conductivity, TDS, salt and temperature.