Pitot Static Tube Set-Up (code EL-FML-10027) measures the point velocity of water in a clear acrylic test section made for 1-inch pipe, using a copper or stainless steel Pitot tube fitted with a vernier scale to set its position across the bore. Readings at different flow rates are used to calibrate the Pitot tube and to plot the velocity profile across the pipe. This Pitot-tube velocity bench is made by JLab in Ambala, India.
Specifications
| Parameter | Detail |
|---|---|
| Pitot tube | Copper or stainless steel, fitted with a vernier scale |
| Test section | Clear acrylic, for 1" dia. pipe |
| Flow measurement | Measuring tank, 40 litres; electronic stopwatch |
| Sump tank | 70 litres |
| Water circulation | FHP centrifugal pump |
| Control panel | On/off switch, mains indicator |
| Utilities | Single phase, 220 V AC, 0.5 kW; water supply and drain |
| Quality check | Each Pitot tube set-up is visually inspected before dispatch |
| Pack quantity | Each (sold individually) |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Product code | EL-FML-10027 |
| Manufacturer | Jain Laboratory Instruments Pvt Ltd (JLab) — OEM |
| Country of origin | India (made in Ambala, Haryana) |
Velocity from the Head Difference
The Pitot opening faces upstream and takes up the stagnation head, while the static holes read the static head of the stream. The difference between them, h in metres of flowing water, is the velocity head, so the ideal point velocity is v = √(2gh). A real tube reads slightly differently, and calibration gives its coefficient C in v = C √(2gh).
Traversing for the Velocity Profile
The vernier scale sets the Pitot tip at measured distances from the pipe axis, so a traverse across the bore gives velocities from wall to wall at each flow rate. Integrating these point velocities over the pipe area and comparing the result with the discharge from the measuring tank calibrates the tube. In laminar flow the profile is a parabola with a centre velocity twice the mean; in turbulent flow, usual at these pump flows, it is much flatter, so the mean comes far closer to the centre value than the one-half found in laminar flow.
Related Flow Measurement and Energy Benches
- Venturimeter and Orificemeter Calibration Set-Up — measures total discharge from a differential head rather than point velocity.
- Study of Flow Measurement Devices — compares a turbine meter, rotameter, orificemeter and venturimeter on one bench.
- Bernoullis Theorem Apparatus — the energy equation behind the Pitot tube formula.
Browse the Fluid Mechanics Lab range for similar benches; it belongs to Fluid Mechanics and Hydraulics Lab Equipment.
Request a Quote
Request a quotation through Add To Cart or the contact page, stating the quantity and the delivery address for the set-up.
Last updated: 30 October 2026