Venturimeter and Orificemeter Calibration Set-Up Manufacturer,Supplier and Exporter in India
Product Code : EL-FML-10022
Jlab Export is a leading Venturimeter and Orificemeter Calibration Set-Up Manufacturer,and suppliers in India
Venturimeter and Orificemeter Calibration Set-Up (code EL-FML-10022) is a water re-circulating bench with two pipelines leaving a common manifold, one carrying a clear acrylic venturimeter and the other an orificemeter, both made for 1-inch pipe. A differential manometer reads the pressure difference across whichever meter is in use, and a measuring tank with stopwatch gives the actual flow. This twin-meter calibration bench is manufactured by JLab in Ambala, India.
Specifications
| Parameter | Detail |
|---|---|
| Meters | Venturimeter and orificemeter, clear acrylic, for 1" dia. pipe |
| Pipe layout | Two pipelines in parallel from a common manifold; downstream valves select the meter and regulate flow |
| Pressure difference | Differential manometer |
| Flow control | Flow control valve and by-pass valve |
| Flow measurement | Measuring tank, 25 litres; electronic stopwatch |
| Sump tank | 50 litres |
| Tank material | Stainless steel |
| Water circulation | FHP centrifugal pump |
| Utilities | Single phase, 220 V AC, 0.5 kW; water supply and drain |
| Quality check | Each calibration set-up is visually inspected before dispatch |
| Pack quantity | Each (sold individually) |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Product code | EL-FML-10022 |
| Manufacturer | Jain Laboratory Instruments Pvt Ltd (JLab) — OEM |
| Country of origin | India (made in Ambala, Haryana) |
Finding the Coefficient of Discharge
At each flow setting the manometer difference is converted to a head h of flowing water, and Bernoulli’s equation with continuity gives the theoretical discharge Qth = a1 a2 √(2gh) / √(a1² − a2²), where a1 is the pipe area and a2 the throat or orifice area. The measuring tank and stopwatch give the actual discharge Qact, and Cd = Qact / Qth. The venturimeter, with its gradual divergent cone, returns a Cd close to 1, while the orificemeter gives a much lower value because of the vena contracta and the eddy losses behind the orifice.
Using the Meters to Measure Flow
Once Cd is known, either meter measures discharge from its manometer reading alone, Q = Cd × Qth. Plotting Qact against √h for each meter gives a straight line whose slope is the meter constant, which is how the second experiment, discharge measurement through each meter, is carried out. Valves on the downstream side switch the flow between the two pipelines and also set its rate.
Related Flow and Energy Benches
- Study of Flow Measurement Devices — adds a turbine flow meter and rotameter to venturi and orifice meters.
- Bernoullis Theorem Apparatus — shows the energy equation that the meter formulas rest on.
- Pitot Static Tube Set-Up — measures point velocity in a pipe instead of total discharge.
Find more benches like it in Fluid Mechanics Lab, a section of Fluid Mechanics and Hydraulics Lab Equipment.