Plasma Extractor Manufacturer,Supplier and Exporter in India
Product Code : EL-PE-2317
Jlab Export is a leading Plasma Extractor Manufacturer,and suppliers in India
After a blood bag has been centrifuged, a plasma extractor squeezes the top layer out into a satellite bag. In this one the main bag sits between a fixed stainless steel plate and a movable transparent plate. A power spring presses the plates together with a clamping force of 20 N, and the frame takes 100, 200, 300 and 400 ml bags.
Specifications
| Plates | Fixed stainless steel plate, movable transparent plate |
| Pressure | Power spring |
| Clamping force | 20 N |
| Bag sizes | 100, 200, 300 and 400 ml |
| External size (W x D x H) | 165 x 250 x 250 mm |
| Net weight | 2.5 kg |
| Gross weight | 2.8 kg |
| Package size (W x D x H) | 310 x 310 x 240 mm |
| Minimum order quantity | 10 units |
| Dispatch time | Usually 4–6 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Lab Exercises in Component Separation
Trainees in a blood bank or MLT lab can practise the whole step with a water-filled practice bag set. They hang the empty satellite bag below the bench edge, place the full main bag against the steel plate and let the spring plate close. Through the clear plate they watch the level fall and clamp the tube when it reaches the line they marked.
They then weigh both bags on the Blood Bank Scale to find how much moved across. Finally they close the tube with the Blood Bag Tube Sealer and check the seal. Teachers use the same exercise to explain why the bag must come out of the centrifuge upright and unshaken.
Frequently Asked Questions
Which bag sizes fit between the plates?
The extractor takes 100, 200, 300 and 400 ml blood bags. Place the bag flat and centred against the steel plate with the outlet tube at the top, so the spring plate presses across its whole face.
How do I know when to stop the flow?
Watch the bag through the transparent plate. As the plasma leaves, the boundary with the red cell layer rises towards the outlet. Clamp the tube before that boundary reaches it, so cells do not pass into the satellite bag.
How should the extractor be cleaned and stored?
Wipe both plates with a mild detergent or the disinfectant your lab uses, then dry them. Avoid corrosive liquids and solvents that cloud the clear plate. Store it with the spring plate closed and released, not held open under load.
What does the 20 N clamping force mean in use?
The spring keeps a steady push on the bag, so the operator does not have to squeeze by hand. Flow is quick at first and slows as the bag empties and the spring extends.
Other Items for the Component Lab
For a quote, use Add To Cart or send us a message through the contact page.
Last updated: 13 November 2026
