Plasma Thawing Bath Manufacturer,Supplier and Exporter in India

Plasma Thawing Bath

Product Code : EL-PTB-2322
Jlab Export is a leading Plasma Thawing Bath Manufacturer,and suppliers in India

Frozen plasma bags lie on a smooth acrylic tray in this tabletop bath while a built-in pump keeps the water moving around them. A micro controller holds the water at the set value, shows it beside the process value and gives an audio-visual alarm if the temperature drifts. The bath thaws at a 37 °C setpoint. A controlled 4 °C mode for cryoprecipitate is offered only as a separate optional configuration and is not used for plasma thawing.

Specifications

TypeTabletop
ControllerMicro controller based
Thawing temperature37 °C setpoint, for quick thawing of fresh frozen plasma
4 °C modeOptional separate configuration for cryoprecipitate (AHF) work, not interchangeable with plasma thawing
DisplaySet value and process value
AlarmAudio-visual, on deviation from the preset temperature
Bag holderSmooth acrylic tray
CirculationIntegrated pump
InsulationFormed in place PUF
OptionDigital indicator-cum-controller with computer data logger
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Training Exercises for Lab Students

MLT and biomedical-engineering students can check the bath against an independent reading. With the pump running, they place the probe of a Calibratable Digital Thermometer at each corner of the tray and in the centre. They record each figure beside the process value and see how the rounded corners and circulation keep the readings close.

For an alarm test they raise the set value a few degrees above the water temperature and note how long the alarm takes to sound, timed with a Digital Stop Watch. They then reset it. Thawing practice uses sealed bags of water frozen in a lab freezer, so students learn to load the tray without touching real plasma.

Frequently Asked Questions

What water level does the tray need?
Fill the tank so the bags on the tray sit fully under water and the pump intake stays covered. A low level leaves part of each bag in air and lets the pump draw air, which upsets the temperature control.

How do I test the deviation alarm?
Change the set value until it differs from the water temperature by more than the alarm allows. The audio-visual alarm should start within a short time. Return the set value, wait for the reading to settle and record the test in the bath log.

What do the set value and process value displays show?
The set value is the temperature you have asked for. The process value is what the sensor reads in the water now. When the two match and stay steady, the bath is ready for loading.

What if a bag leaks in the bath?
Lift the bag out, switch off and drain the tank. Clean the tank and tray, then refill with fresh water before the next use. What happens to the product is decided by the blood bank’s own protocol.