• +91-85699-09696

Plasma Thawing Laboratory Bath Manufacturer,Supplier and Exporter in India

Plasma Thawing Laboratory Bath

Product Code : EL-WB-3981
Jlab Export is a leading Plasma Thawing Laboratory Bath Manufacturer,and suppliers in India

Blood bank laboratories use this tabletop bath to thaw bags of fresh frozen plasma in water at a 37 °C setpoint, close to body temperature. A microcontroller-based controller sets and holds the water temperature. A controlled 4 °C mode for cryoprecipitate (antihaemophilic factor, AHF) work is a separate optional configuration, not part of plasma thawing.

Specifications

TypeTabletop thawing bath
Thawing temperature37 °C setpoint
4 °C modeSeparate optional configuration for cryoprecipitate (AHF) work, not used for plasma thawing
Temperature controlMicrocontroller-based controller
ApplicationThawing fresh frozen plasma (FFP)
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Blood Bank Training Exercises for Students

Medical laboratory technology students learn the daily check that every thawing bath needs. They read the water with a Digital Stem Thermometer at several points, compare it with the controller display and enter both figures in a log sheet. Any gap between the two shows when the bath needs attention.

For practice without blood products, students freeze sealed bags of water and thaw them in the bath, timing each one with a Digital Stop Watch. Comparing a thin, flat bag with a thick one shows why the shape and size of a frozen bag change the thawing time.

Frequently Asked Questions

What temperatures does the bath work at?
It thaws plasma at a 37 °C setpoint. A controlled 4 °C mode for cryoprecipitate work is a separate optional configuration, not interchangeable with the 37 °C thawing protocol.

Why is the thawing water kept at 37 °C and not hotter?
Hotter water would thaw a bag faster, but it can damage the heat-sensitive proteins in plasma. Water near body temperature thaws the bag without overheating the outer layer while the centre is still frozen.

How does a water thawing bath compare with a dry plasma thawer?
In a water bath the warm water touches the whole surface of the bag, so heat passes in quickly and evenly. A dry thawer keeps the bag out of water and warms it through heated plates or air instead.

How should the water temperature be checked?
Place a separate thermometer in the water and compare its reading with the controller display, at least once each working day. Record both readings so that any drift can be seen early.

How do I keep the bath clean?
Empty and refill it with clean water often, wipe the tank and dry it when it is out of use. Clean any spilt plasma at once, so that the water does not become a source of contamination.

Other Laboratory Water Baths

For a quote, use Add To Cart or send us a message through the contact page.