Viscometric Water Bath Manufacturer,Supplier and Exporter in India
Product Code : EL-WB-3980
Jlab Export is a leading Viscometric Water Bath Manufacturer,and suppliers in India
Glass capillary viscometers stand upright in this bath behind a glass window, so the meniscus can be timed between its marks while the water stays at one temperature. A stirrer or circulating pump keeps the water even, and the control holds ±0.1 °C. The bath takes 6, 9 or 12 viscometers and works from ambient +5 °C to 90 °C.
Specifications
| Use | Kinematic viscosity tests with glass capillary viscometers |
| Construction | Double wall, stainless steel inner, powder-coated GI outer |
| Temperature range | Ambient +5 °C to 90 °C |
| Temperature control | ±0.1 °C (±0.01 °C on request) |
| Mixing | Stirrer or circulating pump |
| Heating | Immersion heater with booster for the first heat-up |
| Viewing | Glass window |
| Viscometer holders | 6, 9 or 12, or to suit the application |
| Power supply | 230 V AC, 50 Hz |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Viscosity Practicals for Students
A standard physical chemistry practical finds the viscosity of a liquid with an Ostwald viscometer. Students first time water through the capillary to fix the viscometer constant, then time the test liquid at the same setting with a Digital Stop Watch. Mixtures of Glycerine and water of different strengths show how viscosity climbs with concentration.
In a second exercise the class repeats one liquid at several bath settings and plots the logarithm of viscosity against the inverse of absolute temperature. A Digital Stem Thermometer near the viscometers lets students check the bath reading before each run.
Frequently Asked Questions
How is kinematic viscosity worked out from the flow time?
Multiply the flow time in seconds by the viscometer constant. The constant comes with a calibrated viscometer, or you can find it by timing a liquid of known viscosity, such as water, in the same bath.
Why must the bath temperature stay so steady?
The viscosity of most liquids changes noticeably with each degree, so a small drift shows up in the flow time. This bath holds ±0.1 °C, and ±0.01 °C control can be made on request for closer work.
How many viscometers can be tested together?
The bath is made with holders for 6, 9 or 12 viscometers. A different holder layout can be made to suit the viscometers and samples you use.
How long should a viscometer sit in the bath before timing?
Leave it until the sample has reached the bath temperature. Then take two flow times one after the other. If they agree closely, the sample has settled and the reading can be used.
What does the booster heater do?
It adds extra heat when the bath is first filled with cold water, so it reaches the setting sooner. Once the water is close to the setting, the main heater and the control hold it there.
Other Viscometers and Constant Temperature Baths
- Redwood Viscometer (efflux method for oils)
- Digital Viscometer
- Water Bath Precision
- Refrigerated Water Bath
For a quote, use Add To Cart or send us a message through the contact page.
Last updated: 12 November 2026