Radiant Warmer with Detachable Trolley Manufacturer,Supplier and Exporter in India
Product Code : EL-NP-4020
Jlab Export is a leading Radiant Warmer with Detachable Trolley Manufacturer,and suppliers in India
A radiant warmer holds a heater head over an open baby tray, so staff can reach the infant from every side. On this model the acrylic tray tilts head up or head down and has a foam mattress, and thick acrylic side panels fold down. The unit sits on a detachable trolley, with a 600 W heater and a 60 W observation light.
Specifications
| Baby tray | Acrylic, head up/down tilt, foam mattress |
| Side panels | Thick acrylic, collapsible, access from all sides |
| X-ray | Provision for positioning an X-ray cassette |
| Heater | 600 W |
| Observation light | 60 W bulb |
| Power consumption | 800 W at 100% heater power |
| Supply | 230 V AC ±10%, 50 Hz |
| Trolley | Detachable |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Nursing Skills Lab Exercises
Before a skills session, students fold down each acrylic side panel and raise it again, then tilt the tray head up, head down and back to level. They lay a Newborn Intubation Baby Model on the mattress and practise sliding an X-ray cassette into position without lifting the model.
Biomedical-engineering students measure the observation light at mattress level with a Digital Lux Meter and test the mains lead for earth continuity with a Digital Multimeter. At the end, two students lift the warmer off its trolley, wipe both parts and store them with the side panels closed.
Frequently Asked Questions
What does 800 W at 100% heater power mean?
It is the most power the unit draws, with the heater at full output. Plan the 230 V socket and circuit for that load, and avoid sharing it with other heavy equipment.
How are the side panels used?
The thick acrylic panels fold down to open the tray for care from any side, and fold back up to close it in. Check that each panel is fully raised and secure before staff step away from the warmer.
How does the head up/down tilt work?
The acrylic tray tilts so the head end sits higher or lower than the feet. Set the tilt with the tray empty and check it is level again before the next use. Clinical positioning follows the hospital’s own protocol.
Why does the trolley detach?
The warmer can be separated from its trolley to pass through narrow doors or to be stored, and each part is lighter to carry. Refit it and check it is firmly seated before use.
When should the foam mattress be replaced?
Replace it when the cover is torn, cracked or stained in a way that will not clean, because liquid can then soak into the foam. Keep a spare so the warmer is not out of use while a new mattress is ordered.
Other Neonatal Care Equipment
For a quote, use Add To Cart or send us a message through the contact page.