Rectangular Muffle Furnace Laboratory Model Manufacturer,Supplier and Exporter in India
Product Code : EL-MF-2061
Jlab Export is a leading Rectangular Muffle Furnace Laboratory Model Manufacturer,and suppliers in India
A muffle furnace heats samples inside a closed refractory chamber, with the heating wire set in grooves round it, so the charge never touches the elements. This rectangular laboratory model reaches 1000 °C and works at up to 940 °C under an energy regulator. It is made in eight inner muffle sizes, from 225 x 100 x 100 mm to 450 x 225 x 225 mm.
Specifications
| Outer shell | Thick mild steel, stoving enamel finish |
| Heating elements | Wire, held in grooved refractory, replaceable |
| Maximum temperature | 1000 °C |
| Working temperature | 940 °C |
| Standard control | Energy regulator with pyrometer |
| Option | Digital temperature indicator-cum-controller |
| Inner muffle sizes (mm) | 225 x 100 x 100, 250 x 125 x 125, 300 x 125 x 125, 300 x 150 x 150, 300 x 200 x 200, 350 x 175 x 125, 475 x 175 x 125, 450 x 225 x 225 |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Ash Content Practicals for Students
Food science and biochemistry students find the ash content of flour, dried leaves or another plant sample. They weigh the sample into a Porcelain Crucible with Lid, char it gently over a flame, then burn it to a white or grey ash in the furnace. The crucible is lifted out with Crucible Tongs.
After each heating the crucible cools in a Desiccator Plain before weighing, so it does not pick up moisture from the air. Students repeat heating, cooling and weighing until two weights agree, which shows constant mass. The ash as a percentage of the dry sample is then worked out.
Frequently Asked Questions
Should I choose the energy regulator or the digital controller?
The energy regulator sets the heater to a share of full power, and the pyrometer shows the temperature it reaches. The digital indicator-cum-controller holds a temperature you set and shows it, which suits repeated ashing at one fixed temperature.
Why is the working temperature lower than the maximum?
The furnace can reach 1000 °C, but long runs are meant for 940 °C or below. Working under the maximum is easier on the wire elements and the refractory, so they last longer.
Can the heating elements be replaced?
Yes. The wire sits in grooves in the refractory and can be taken out and replaced when it fails. Switch off, let the furnace cool fully and have the new element fitted by someone trained in the work.
Which inner muffle size should I order?
Count the crucibles or dishes you heat at one time and leave space between them for even heating. The 225 x 100 x 100 mm chamber suits a few small crucibles, while the deeper and wider sizes take larger batches or bigger parts.
Why should a new furnace be heated slowly the first time?
New refractory holds some moisture. Raising the temperature in stages lets that moisture escape gently and reduces the risk of cracking the muffle.