• +91-85699-09696

Temperature and Precision Control Water Bath (Hospital Lab) Manufacturer,Supplier and Exporter in India

Temperature and Precision Control Water Bath (Hospital Lab)

Product Code : EL-WB-2281
Jlab Export is a leading Temperature and Precision Control Water Bath (Hospital Lab) Manufacturer,and suppliers in India

Built for work that needs the same temperature in every corner, this water bath has a solid state controller and a stirrer that keeps the water moving. The rod and blade are stainless steel, as is the 430 x 240 x 265 mm chamber. It covers 50 °C above ambient up to 100 °C.

Specifications

ControlSolid state temperature controller
StirrerStainless steel rod and blade
Temperature range50 °C above ambient to 100 °C
ChamberStainless steel sheet
Chamber size430 × 240 × 265 mm
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Physical Chemistry Experiments at Set Temperatures

Viscosity changes quickly with temperature, so a glycerol and water mixture makes a clear practical. Students hold the sample beaker in the stirred bath and confirm its temperature with a Digital Stem Thermometer. They then take a reading on a rotational Viscometer and repeat at three or four settings to plot viscosity against temperature.

For a solubility curve, a saturated solution of benzoic acid in water stands in the bath at each setting until it settles. A filtered sample is then titrated with alkali from a Burette, Borosilicate Glass, with PTFE Stopcock to find how much acid dissolved.

Frequently Asked Questions

How can I check that the temperature is even across the chamber?
With the stirrer running, move one thermometer to the corners, the centre and two depths, and let it settle at each point. The readings should agree closely. If one spot runs warm or cool, a vessel may be blocking the flow from the blade.

Why does the bath overshoot the setting when it first heats up?
The heater is still hot when the controller cuts it, so the water climbs a little past the setting before it falls back. Set slightly low at first, then raise the setting in small steps as the water nears the target.

How deep should a flask sit in the water?
Keep the liquid inside the flask below the bath’s water level, so all of it is surrounded by water at the set temperature. The 265 mm deep chamber takes tall flasks and boiling tubes clamped upright.

Why does the bath take a while to reach temperature?
A full chamber holds a large volume of water, and all of it has to warm up. The same volume makes the bath steady once it gets there, as a cold flask put in barely moves the water temperature.