Vertical Sterilizer Manufacturer,Supplier and Exporter in India
Product Code : EL-MS-3760
Jlab Export is a leading Vertical Sterilizer Manufacturer,and suppliers in India
Hospitals use a vertical sterilizer to steam dressing drums, linen packs and instruments in an upright chamber surrounded by a jacket. The standard model works at 1.26 kgf/cm² and 121 °C. A high-speed model works at 2.2 kgf/cm² and 134 °C with a shorter hold and exhaust. Both run on a 400/440 V three-phase supply.
Specifications
| Type | Vertical steam sterilizer with jacket |
| Steam working pressure | 1.26 kgf/cm² (2.2 kgf/cm² on the high-speed model) |
| Steam working temperature | 121 °C (134 °C on the high-speed model) |
| Hydraulic test | Jacket at twice the working pressure, chamber at one and a half times |
| Sterilization period | 20–25 minutes (5–7 minutes on the high-speed model) |
| Steam exhaust | 5–7 minutes (1 minute on the high-speed model) |
| Operating voltage | 400/440 V, 3 phase, AC, 50 Hz |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Training Practicals in the CSSD
Nursing and CSSD trainees learn the loading pattern first. They stand a packed Dressing Drum and a Sterilizer Shallow Dressing Drum upright in the basket with their belts open, leaving gaps so steam can flow between them. The tutor then runs a cycle while students note the time the gauge reaches working pressure, the hold time and the exhaust time in a log book.
In a second exercise, students compare the standard and high-speed settings on paper and explain why the shorter cycle needs the higher pressure. Loaded drums are moved out on an Instrument Trolley once the gauge has fallen to zero and the lid is safe to open.
Frequently Asked Questions
What is the high-speed version?
It is the same vertical design run at 2.2 kgf/cm² and 134 °C. The period drops to 5 to 7 minutes and the steam exhaust to 1 minute. Which cycle is used for which load is decided by the hospital protocol and trained CSSD staff.
What do the hydraulic test figures mean?
During manufacture the jacket is tested with water at twice the working pressure and the chamber at one and a half times. A water test at a higher pressure than the unit works at shows that the shells and joints hold without leaking.
Why does the pressure not reach the working value?
The usual causes are a lid that is not sealed evenly, a worn gasket, too little water, a heater that has failed or a valve left open. Switch off and let the unit cool before any check, and call a technician if the heater is suspected.
Why is there water inside the pressure gauge?
Condensed steam can work its way into the gauge, and the dial then mists over or reads wrongly. Stop using the gauge for readings and have a technician drain or replace it before the next run.
Can it run on a single-phase supply?
No. This model is built for a 400/440 V three-phase AC supply at 50 Hz, so the site needs a three-phase point. The Horizontal Cylindrical Sterilizer is the model in this range made for a single-phase supply.