Digital Infrared Moisture System Manufacturer,Supplier and Exporter in India
Product Code : EL-MM-1259
Jlab Export is a leading Digital Infrared Moisture System Manufacturer,and suppliers in India
Infrared rods, rather than a heat lamp, warm the sample in this dry-and-weigh moisture system, and built-in software controls how fast the heat rises. It weighs samples of 10 to 100 g to 0.01 g as they dry. Temperature is set from ambient to 250 °C with a platinum sensor, and the result comes out on a wet or dry basis.
Technical Details
| Measurement method | Infrared heating with detection of weight loss |
| Heat source | Infrared rods |
| Temperature setting | Ambient to 250 °C |
| Temperature control | Platinum temperature sensor |
| Accuracy | ± 0.1% |
| Sample weight | 10 g to 100 g (optional up to 300 g) |
| Sensitivity | 0.01 g |
| Working range | 0.1% to 99.9% |
| End of analysis | Time-out, 1 to 999 min |
| Result readability | 0.10%, moisture in per cent, wet or dry basis |
| Power and accessories | 230 V ± 10% AC 50 Hz, 200 W; aluminium disposable pans, tweezers, mains lead, manual; printer optional |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Loss on Drying Practicals in Pharmaceutics
In a tablet granulation practical, students pass the wet mass through Granulating Sieves and spread the granules in a Tray Drying Oven. Every quarter of an hour they take a few grams out and find the moisture on this system. The curve of moisture against drying time shows when the granules are dry enough to compress.
Food and agriculture students can run one sample at two drying temperatures. A setting that is too hot reads high once the material starts to break down.
Frequently Asked Questions
Should I report moisture on a wet basis or a dry basis?
Wet basis is the loss in mass divided by the starting mass, and it is the usual figure for food, grain and loss on drying. Dry basis divides the loss by the dried mass. The software gives either, so state which one you report.
How do I choose the drying temperature?
Pick the highest setting at which the sample loses water without browning, melting or giving off fumes. Test it once against the oven method on the same material, then keep that temperature as the method for that product.
What does the time-out end mode do?
You set a drying time from 1 to 999 minutes. The system heats and weighs for that time and then shows the final moisture. Use the same time for every sample of one product so results can be compared.
Why use the disposable aluminium pans and tweezers?
A fresh pan carries no residue from the last sample, so its tare stays clean. Handling pans with the tweezers keeps moisture and grease from fingers off the pan, which matters when the system weighs to 0.01 g.
Other Moisture Analysers
- Infra Red Moisture Balance (infrared bulb, moisture scale)
- Moisture Analyzer
- Halogen Moisture Balance
For a quote, use Add To Cart or send us a message through the contact page.