Interferential Therapy Manufacturer,Supplier and Exporter in India
Product Code : EL-PE-3844
Jlab Export is a leading Interferential Therapy Manufacturer,and suppliers in India
Two carrier frequencies, 2 kHz and 4 kHz, are available on this interferential unit. The base frequency is adjustable from 0 to 150 Hz and the sweep from 0 to 100 Hz. It works in 4 pole interferential, 4 pole vector and 2 pole modulated modes. Separate displays show the intensity of each channel, and a digital timer with alarm cuts the output when it reaches zero.
Specifications
| Medium (carrier) frequency | 2 kHz and 4 kHz |
| Base frequency | 0 to 150 Hz |
| Sweep frequency | 0 to 100 Hz |
| Therapy modes | 4 pole interferential, 4 pole vector, 2 pole modulated |
| Timer | Digital, with alarm; output cuts off at zero |
| Intensity display | CH-1 digital, CH-2 LED |
| Other controls | Sweep program, balance control of intensity |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Lab Exercises on Carrier and Beat Frequency
Biomedical and physiotherapy students connect each channel to a resistor load and display both on a Dual Trace Oscilloscope. They measure the period of the 2 kHz and 4 kHz carriers on the screen. Then they switch to 2 pole modulated output and read the beat envelope against the base frequency setting. Turning on the sweep shows the envelope speeding up and slowing down.
A timing exercise follows. Students set the timer, start a Digital Stopwatch with the unit and confirm that the alarm sounds and the output stops when the display reaches zero.
Frequently Asked Questions
Should we use the 2 kHz or the 4 kHz carrier?
Both carry the same beat frequency. A higher carrier meets lower skin impedance, so the current passes the skin with less sensation, while the lower one is felt more strongly. The physiotherapist chooses the carrier for each treatment.
How is the base frequency produced?
The two channels run at slightly different medium frequencies. Where they cross, the currents add and cancel in turn, and the rate of that rise and fall is the beat, set here from 0 to 150 Hz. Sweep moves the beat up and down within a 0 to 100 Hz span.
Why does the stimulation feel weak or stop during use?
Check that the timer has not run out, then look at the electrode contact, the dampness of sponge covers and the leads for breaks. A pad lifting at one edge or a dry cover cuts the current. If both displays show output but nothing is felt, have the unit serviced.
Why do the two channels have different displays?
Channel 1 shows its intensity on a digital display and channel 2 on an LED display, so each circuit can be read at a glance. Watch both when using the balance control, which moves intensity from one channel to the other.
Other Electrotherapy Units
- Interferential Therapy Machine
- Diagnostic Muscle Stimulator
- Digital Ultrasonic Physiotherapy Equipment
For a quote, use Add To Cart or send us a message through the contact page.
Last updated: 13 November 2026