Perchloric Fume Hoods Manufacturer,Supplier and Exporter in India
Product Code : EL-LFH-3875
Jlab Export is a leading Perchloric Fume Hoods Manufacturer,and suppliers in India
Perchloric acid fumes leave perchlorate salts behind wherever they condense, and those deposits have to be washed away. This hood has a wash-down system behind the baffles, fed by a water pump that keeps the pressure constant. The inner chamber is stainless steel with coved corners, set in a tri-wall body with a chain and sprocket sash.
Specifications
| Construction | Tri-wall |
| Inner chamber | Stainless steel with coved corners, grade 316 as an option |
| Wash-down | Spray system behind the baffles, with water pump for constant pressure |
| Sash support | Chain and sprocket |
| Sash stop | 457 mm (18 in); a higher sash falls back to this level unless held |
| Sash key | Disables the creep-down and locks the sash fully closed |
| Sidewall airflow | Shaping vanes to increase the sweep at the sidewalls |
| Lighting | Pre-wired, electronically ballasted, instant start; over 1076 lux on the work surface in zero ambient light |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
College Practicals in a Perchloric Hood
Wet digestion of plant, soil or food samples with nitric and perchloric acids is a standard step in agricultural, environmental and food chemistry labs. The samples are heated in the flasks of Kjeldahl Digestion Units set inside the hood. The clear digest is then diluted and analysed, for example for phosphorus on a Photo Electric Colorimeter.
At the end of each session students switch on the wash-down to rinse the baffles, then wipe the worktop. Before any heating starts, a quick smoke test at the sash opening shows the air being drawn steadily inward.
Frequently Asked Questions
Can I use perchloric acid in an ordinary fume hood?
Heated perchloric acid should not go in an ordinary hood. Its fumes leave perchlorate salts on the baffles and in the duct, and an ordinary hood has no way to wash them out. This hood has a wash-down system behind the baffles for that reason.
How often should the wash-down be run?
Run it after each session of work with heated perchloric acid. Keep a log of when and for how long it ran, as your lab rules require. The water pump keeps the wash-down pressure steady each time.
Why are the inner corners coved?
Rounded corners leave no sharp seams where salts can gather, so wash water and wiping reach every part of the chamber. The stainless steel chamber can be upgraded to grade 316.
Why does the sash slide back down when I let go of it?
The sash stop sets the working opening at 457 mm (18 in). You can raise the sash fully to set up apparatus, but it returns to that height unless held. Only a key holder can switch this off.
What should be kept out of a perchloric acid hood?
Keep organic solvents, oils, paper, wood and other combustible materials out of it, and do not store chemicals inside. Perchloric acid reacts with organic matter, and anything left in the chamber collects perchlorate deposits.
Other Fume Hoods and Exhaust Parts
For a quote, use Add To Cart or send us a message through the contact page.
Last updated: 12 November 2026