Shortwave Medical Diathermy Manufacturer,Supplier and Exporter in India

Shortwave Medical Diathermy

Product Code : EL-PE-3823
Jlab Export is a leading Shortwave Medical Diathermy Manufacturer,and suppliers in India

Six filament voltage steps and a 30 minute timer are the main controls on this shortwave medical diathermy. It is a high-frequency generator working at 27.12 MHz, with a maximum output power of 500 W. When the timer runs out, the unit switches off its RF power by itself. Treatment with it is planned and given by a physiotherapist.

Technical Details

Operating frequency27.12 MHz
Maximum output power500 W
Filament voltageAdjustable in six steps
Timer30 minute treatment timer
End of timerRF power switches off
Minimum order quantity1 unit
Dispatch timeUsually 14–18 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Electrotherapy and Biomedical Lab Practicals

With the tutor in charge, students run the unit into a test load instead of a person. A plastic Beaker of water stands between the electrodes, and students bring up the output slowly and feel the water warm over a few minutes. Holding the neon bulb of a Neon Tester near the electrodes shows the field: the bulb glows when output is present. Watches, keys, phones and other metal or electronic items stay well away from the field.

In the biomedical-engineering lab, the unit is unplugged and an Insulation Tester is used to check the mains lead. Students also check the earth path, look over the plug and cables, and confirm the timer cuts the RF output at zero.

Frequently Asked Questions

What does the filament voltage control do?
It sets the heating voltage for the filament of the generator valve in six steps. Matching it to the local mains supply lets the valve run at its proper heat, which helps it give steady output and last longer.

What happens when the timer runs out?
The unit switches off the RF power automatically at the end of the set time, so output does not continue if staff are busy elsewhere. The timer runs up to 30 minutes, and the physiotherapist decides the setting.

Why does the unit cut out after it has warmed up?
A unit that stops when hot and works again once cool often has blocked air vents or a failing part near the valve or transformer. Leave space around the vents, and have a biomedical engineer test it before further use.

Can the unit run from an extension lead?
It should go straight into an earthed wall socket. Extension leads and loose sockets can drop the supply or break the earth, and a damaged mains cable may fail only when flexed. Check the plug, fuse and cable entry from time to time.