Ampoule Filling & Sealing Device

Product Code : EL-PLE-2231
Jlab Export is a leading Ampoule Filling & Sealing Device Manufacturer,and suppliers in India

This bench device fills glass ampoules and then seals them. An electrically driven syringe piston delivers the liquid, and its stroke is adjustable to set the quantity filled into each ampoule. The burner supplied with it melts the ampoule neck closed once it is filled.

Specifications

OperationElectrically operated
FillingSyringe piston with adjustment for the quantity of liquid filled
SealingBurner supplied for sealing the ampoules
Minimum order quantity1 unit
Dispatch timeUsually 14–18 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Parenteral Practicals with the Device

In the B.Pharm sterile products practical, students prepare a simple injection such as sodium chloride solution and fill it into ampoules. The empty ampoules are first rinsed on the Ampoule Washing Device and dried. After filling, each neck is closed by tip sealing or pull sealing in the burner flame.

The sealed ampoules are then sterilised in a Laboratory Autoclave and checked for leakers by immersing them in a dye solution under vacuum. Students count the ampoules that take up dye and relate the failures to their sealing technique.

Frequently Asked Questions

What is the difference between tip sealing and pull sealing?
In tip sealing the very end of the neck is melted until the glass runs together into a bead. In pull sealing the neck is heated lower down while the ampoule turns, and the tip is drawn off so the glass closes in a twisted seal. Pull sealing is slower but gives a more secure closure.

Why do some sealed ampoules leak?
Leakers usually come from a fine capillary left open at the tip, or a crack at the seal from uneven heating. Turn the ampoule steadily in the flame and check every sealed batch with a dye leaker test.

Why does a bubble or a blob form at the seal?
Holding the neck in one spot too long overheats the glass, so it balloons out or gathers into a thick lump that can crack as it cools. Keep the ampoule turning and heat only the neck.

How is the fill volume set?
Adjust the stroke of the syringe piston, fill a few trial ampoules and check their contents in a small measuring cylinder. Correct the setting before filling the main batch.

Why must the neck stay dry during filling?
Liquid left on the neck boils or chars in the flame and spoils the seal. Let the needle reach well into the ampoule and withdraw it without touching the neck.

Other Filling and Sealing Equipment

For a quote, use Add To Cart or send us a message through the contact page.

Last updated: 12 November 2026


   