Sodium and potassium in serum, urine, water and soil extracts are read on this digital flame photometer, and calcium and lithium can be added with optional filters. A separate oil-free compressor unit feeds air to the nebulizer, which sprays the sample into an LPG flame. A silicon photodiode reads the light passed by a narrow-band filter, and the LED display shows ppm or meq.
Specifications
| Elements | Na and K; Ca and Li optional |
| Range | Na and K 0–100 ppm; Ca 15–100 ppm; Li 10–100 ppm |
| Serum range | Na 0–200 meq/l; K 0–10 meq/l |
| Sensitivity | Na and K 5 ppm; Ca and Li 10 ppm |
| Measurement | One element at a time, or Na and K together on the two-channel clinical version |
| Readout | 2½ digit 7-segment LED, dual on the clinical version |
| Filters | Narrow-band interference glass filters on a thumb-turned disc |
| Flame system | LPG and dry oil-free air, electronic ignition by switch |
| Warm-up time | 10 minutes |
| Units | Main unit 375 x 245 x 220 mm, 7.5 kg; compressor unit 290 x 255 x 210 mm, 8 kg; 230 V AC, 50 Hz |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Experiments for Clinical and Soil Science Students
A common clinical practical is the estimation of serum sodium and potassium. Students weigh dried Sodium Chloride and Potassium Chloride on an Electronic Balance Analytical to make stock standards and dilute them into a working series. Serum spun off in a Medico Centrifuge Machine is diluted the same way. On the two-channel version, sodium and potassium are read together from one aspiration.
In soil science classes, students read potassium in soil extracts and compare samples from different plots. A good teaching exercise is to plot the standard readings by hand on graph paper and read each sample off the curve, which shows how concentration is worked out from emitted light.
Frequently Asked Questions
How long should the instrument warm up?
Allow about 10 minutes after lighting the flame. In that time the burner and detector settle and the readings stop drifting. Set the zero with distilled water and the top standard only after the warm-up is over.
How do I change from sodium to potassium?
The filters sit on a disc inside the main unit. Turn the disc with your thumb to bring the filter for the element you need into the light path. Then reset the zero and the standard for that element before reading samples.
How do the single-channel and two-channel versions differ?
The single-channel version reads one element at a time on one LED display. The two-channel clinical version has two displays and shows sodium and potassium together from one aspiration, which saves time on serum and urine batches.
What does the compressor unit do?
It supplies oil-free compressed air at a steady pressure set by its regulator. The air stream draws the sample into the nebulizer and sprays it into the flame, so a steady airflow gives a steady flame and steady readings.
Why do readings drift during a run?
Drift usually comes from changes in gas or air pressure, a part-blocked nebulizer or a sample spraying unevenly. Check a standard every few samples, flush with distilled water between readings and clean the nebulizer if the sample uptake slows.