Digital HB Reader LCD Alpha Numeric Manufacturer,Supplier and Exporter in India

Digital HB Reader LCD Alpha Numeric

Product Code : EL-PALE-9840
Jlab Export is a leading Digital HB Reader LCD Alpha Numeric Manufacturer,and suppliers in India

The HB reader is a small photometer set up for one test: haemoglobin by the cyanmethaemoglobin method. Green light from an LED at 546 nm passes through the diluted sample in a square plastic tube, and a silicon photodiode measures it. A microcontroller converts the reading and shows the result in g/dl on a one-line alphanumeric display, with no calculation needed.

Specifications

MethodCyanmethaemoglobin
Wavelength546 ± 5 nm
Light sourceGreen LED
DetectorSilicon photodiode
Display16 x 1 alphanumeric LCD, result in g/dl
ModesMeasurement and Calibration, auto zero
Default standard value15 g/dl, can be changed in Calibration mode
Memory40 tests
Keypad2 x 2 soft touch
Sample tubesSquare 10 x 10 mm, plastic; minimum volume 1 ml
PowerMains 220 V ± 10%, 50 Hz, single phase; 12 V DC battery
Size and weight215 x 150 x 100 mm, 1.5 kg net
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Haemoglobin Practicals for Teaching Labs

A common class exercise compares two methods on the same sample. Students add a measured drop of blood to the cyanide reagent with a Variable Volume Micropipette, using a fresh tip from the Micropipette Tips pack for each sample. After mixing and standing, the solution is poured into the square tube and read. The same blood is then tested by the acid haematin method on a Haemoglobinometer, and the class discusses why the visual result differs.

A second exercise covers calibration. Students read a standard solution and check that the stored value of 15 g/dl matches its label. If a different standard is used, they enter its value in Calibration mode.

Frequently Asked Questions

Why must the diluted sample stand before it is read?
The reagent first converts haemoglobin to methaemoglobin and then to cyanmethaemoglobin, and this takes a few minutes. Reading too early catches the reaction part way, so the result comes out low. Follow the standing time given with your reagent kit.

How is the reader calibrated?
Put a cyanmethaemoglobin standard in a clean tube and read it in Calibration mode. The reader stores 15 g/dl as the standard value by default. If your standard has a different value, enter that value, then return to Measurement mode for samples.

Does the reagent need special handling?
Yes. Cyanmethaemoglobin reagent contains a small amount of cyanide salt. Keep it labelled and capped, never pipette it by mouth, and dispose of used solutions as your laboratory rules require. Wash hands after the practical.

How does this differ from the Sahli method?
In the Sahli method the observer matches a brown colour by eye, so two people can read the same tube differently. The reader measures the light that passes at 546 nm with a photodiode and shows the figure itself, which removes the colour-matching step.

What does the memory store, and can it run without mains?
It keeps the last 40 test results, so a batch can be noted down at the end of the class. It runs on 220 V mains and also takes a 12 V DC battery supply where mains is not at hand.