Haemoglobin Meter

Product Code : EL-BTI-2158
Jlab Export is a leading Haemoglobin Meter Manufacturer,and suppliers in India

This meter measures the haemoglobin concentration of a diluted blood sample and shows it directly in Hb units. It uses the cyanmethaemoglobin method: green light at 546 nm from an LED passes through the sample to a silicon photodiode. It reads 0 to 30 g/dl from a 1 ml sample, sets zero and calibrates automatically, and keeps the factor in memory.

Specifications

Measuring methodCyanmethaemoglobin
Measuring range0–30 g/dl Hb
Wavelength546 nm, green LED light
Sample volume1 ml
Display3 digit, 7-segment red LED, direct Hb units
Keyboard3 soft-touch membrane keys
Zero setting and calibrationAutomatic, with the factor stored in memory
DetectorSilicon photodiode
Power, size and weight230 V ±10% AC, 50 Hz; 195 x 235 x 105 mm; about 2.5 kg
Supplied withSet of 5 matched test tubes, spare lamp, operation manual, dust cover
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Haematology Practicals for MLT Students

Haemoglobin estimation is one of the first haematology practicals. Students take a small volume of blood with a Micropipette, wipe the outside of the tip and mix the blood into cyanmethaemoglobin reagent. Once the colour has fully developed, the tube is read against a reagent blank and the result appears in Hb units.

Teachers often run the same blood by Sahli’s acid haematin method on a Haemoglobinometer, where the colour is matched by eye against standard strips. Comparing the two shows students how a photometric reading at a fixed wavelength takes the guesswork out of colour matching.

Frequently Asked Questions

How does the cyanmethaemoglobin method work?
The reagent turns the haemoglobin in the sample into cyanmethaemoglobin, a stable coloured compound. Its absorbance at 546 nm rises in proportion to haemoglobin concentration, so the meter can show the result directly. The reagent contains cyanide compounds, so follow your lab rules for handling and waste.

What range does it read, and how often must I calibrate?
It reads 0 to 30 g/dl on a 3-digit red LED. Zero setting and calibration are automatic, and the calibration factor stays in memory between runs. Recalibrate with a haemoglobin standard when you start a new lot of reagent.

Why do student results come out too high or too low?
A common cause is a wrong blood volume, from air bubbles or blood left on the outside of the pipette tip. Poor mixing and reading before the colour has fully developed also give errors. Cloudy samples read falsely high, and scratched or dirty tubes add to the absorbance.

Can I use any test tube?
Use the matched test tubes from the set of five supplied with the meter, so every reading passes through the same path length. Fill each tube with the diluted sample, wipe the outside and replace any tube that is scratched.

How should the meter be looked after?
Keep the dust cover on when the meter is idle. Wipe any spill from the tube holder straight away, because reagent drying inside the holder clouds the light path. It runs on 230 V AC, 50 Hz mains.