Balanced Cell Photo Colorimeter Manufacturer,Supplier and Exporter in India
Product Code : EL-BTI-2155
Jlab Export is a leading Balanced Cell Photo Colorimeter Manufacturer,and suppliers in India
A balanced-cell colorimeter compares the light passing through the sample with a reference cell and shows the result on a 0 to 1000 absorbance scale. The scale rises in step with concentration, so a sample can be read directly against a standard solution. Blue, green and red filters are fitted, and each reading needs less than 4 ml in a 15 x 62 mm tube.
Specifications
| Scale | 0–1000 (absorbance scale) |
| Display | 3 digit, 7-segment bright red LED |
| Filters | Blue 440 nm, green 540 nm, red 660 nm on a rotating disc |
| Test tube | Standard size, 15 x 62 mm |
| Sample quantity | Less than 4 ml |
| Light source | 6.3 V, 300 mA |
| Amplifier | Solid state, IC based |
| Power | 230 V ±10% AC, 50 Hz |
| Dimensions | 245 x 300 x 205 mm (approx.) |
| Supplied with | Matched test tubes, operation manual, dust cover |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Soil and Fertiliser Lab Exercises
Soil testing labs use it for available phosphorus. Students shake air-dried soil with an extracting solution, filter it and develop the molybdenum blue colour with a reagent made from Ammonium Molybdate. The blue extract is read on the red 660 nm filter against a set of phosphate standards. Organic carbon by the colorimetric dichromate method is read on the same red filter.
In a college chemistry class the instrument shows Beer’s law with a blue solution. Students make Copper Sulphate standards with Bulb Pipettes, Class B Soda Glass, One Mark JLab. They read the set on the red filter and check that the scale reading doubles when the concentration doubles.
Frequently Asked Questions
What do the 0 to 1000 readings mean?
They are absorbance units on a scale that rises in proportion to concentration. Read a standard and the sample on the same filter. Sample concentration then equals the standard concentration times the sample reading, divided by the standard reading. For a spread of concentrations, plot a line from several standards instead.
Which filter should I use for soil tests?
Use red, 660 nm, for blue and green colours such as molybdenum blue phosphorus and dichromate organic carbon. Green, 540 nm, suits pink, red and purple colours. Blue, 440 nm, suits yellow colours such as the vanadomolybdate method for phosphorus.
How much sample does each reading need?
Less than 4 ml, in standard 15 x 62 mm test tubes. Use the matched tubes supplied with the colorimeter and fill each one to the same height. Wipe the outside so dust or fingerprints do not add to the reading.
How does it differ from the Digital Photo Colorimeter?
This model has three filters and a 0 to 1000 absorbance scale on a red LED display, which suits routine soil and fertiliser tests. The Digital Photo Colorimeter has more filters and reads optical density directly from a smaller sample.
Should I recheck the standard during a long batch?
Yes. Set the blank and read one standard at the start of each session and again after any break. If the standard reading has moved, set the blank again before you read further samples.
Other Colorimeters for Testing Labs
- Microprocessor Photo Colorimeter (eight filters, K factor memory)
- Digital Photoelectric Colorimeter
- Photo Electric Colorimeter
- Auto Colorimeter
For a quote, use Add To Cart or send us a message through the contact page.