Digital Conductivity Meter

Product Code : EL-CM-2192
Jlab Export is a leading Digital Conductivity Meter Manufacturer,and suppliers in India

This microcontroller conductivity meter comes in versions that also read temperature and correct for it automatically, and in a conductivity-only version with manual compensation. Every version has a 3½ digit LED display, a platinum dip-type cell and automatic ranging up to 1000 mS/cm. All of them offer automatic cell constant measurement and a Check facility for calibrating the instrument.

Specifications

MeasuresConductivity and temperature, or conductivity only, by version
Display3½ digit LED
Conductivity ranges0–200 µS/cm, 0–2, 0–20, 0–200 and 0–1000 mS/cm, automatic in 5 or 6 ranges by version
AccuracyConductivity ±0.5% FS ±1 digit, temperature ±0.1 °C ±1 digit
Resolution0.1 µS/cm, temperature 0.1 °C
Temperature compensationAuto and manual, 0–50 °C on temperature versions, manual 0–50 °C on the conductivity-only version
Cell constant and calibrationAdjustable on the digital display, automatic cell constant measurement, Check facility
Measuring cellPlatinum dip type
Power, size and weight12 V DC adapter for 110–250 V AC, 50 Hz, 270 x 202 x 80 mm, about 0.8 kg
Supplied withConductivity cell, temperature probe, cell stand, dust cover, operation manual
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Experiments on Conductance and Temperature

With a temperature version, students can see how strongly conductivity depends on temperature. They dissolve Sodium Chloride in warm water and read the solution as it cools on the bench, once with ATC switched off and once with it on. Without compensation the reading falls steadily as the solution cools. With compensation it stays almost level.

A second practical compares a strong acid with a weak one. Students titrate Hydrochloric Acid Dilute, then a dilute acetic acid solution made from Glacial Acetic Acid, against sodium hydroxide while a Magnetic Stirrer keeps the beaker mixed. The strong acid gives a V-shaped graph. The weak acid gives a line that rises gently, then more steeply after the end point.

Frequently Asked Questions

What is the Check facility used for?
The Check facility is provided to calibrate the instrument before you start measuring, so run it at the start of each session. After that, confirm the cell constant with a standard solution of known conductivity, using the automatic cell constant measurement.

Should I choose automatic or manual temperature compensation?
Choose a version with automatic compensation if samples reach you at different temperatures, such as field samples or titrations that warm up. Its temperature probe reads the sample and the meter corrects between 0 and 50 °C. The conductivity-only version suits work at a steady room temperature, where you set the temperature by hand.

How close is the temperature reading?
On the versions that measure temperature, the reading is within ±0.1 °C ±1 digit, with 0.1 °C resolution. Conductivity accuracy is ±0.5% of full scale ±1 digit on every version, on all ranges up to 1000 mS/cm.

Why must the cell be dipped to the right depth?
The cell constant assumes the sample fills the space between the platinum plates. If the liquid is too shallow or air is trapped, the reading falls. Dip the cell deep enough to cover the plates, tap it to release bubbles and keep it clear of the beaker wall and bottom.

Close Models and Other Conductivity Instruments