Dissolution Rate Test Equipment Manufacturer,Supplier and Exporter in India
Product Code : EL-PLE-2250
Jlab Export is a leading Dissolution Rate Test Equipment Manufacturer,and suppliers in India
A dissolution test measures how fast the drug in a tablet passes into solution, which a disintegration test alone cannot show. In this equipment the tablet sits in a stainless steel basket that turns at 25 to 150 rpm inside a 1000 ml cylindrical vessel. Moving water around the vessel holds 37 °C ± 0.5 °C throughout the test. It comes as a single unit or as combined units for 3 or 6 tests.
Specifications
| Vessel | 1000 ml cylindrical vessel with flanges near the top |
| Cover | 4-part cover |
| Sample holder | Stainless steel basket |
| Speed | 25 to 150 rpm, set by a speed controller |
| Bath | Moving water held at 37 °C ± 0.5 °C during the test |
| Versions | Single test, combined unit for 3 tests, combined unit for 6 tests |
| Power | 220 V AC, 50 Hz, single phase |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Drug Release Practicals
A standard practical is a release profile of a plain tablet such as paracetamol. Students withdraw samples at set times, filter them and read the absorbance on a Double Beam UV-VIS Spectrophotometer against a standard curve. They then plot the percentage released against time.
A second exercise compares plain tablets with tablets coated in a Tablet Coating Pan, or tablets pressed at two different settings. On the 3-test and 6-test units, several batches can run side by side, which saves class time.
Frequently Asked Questions
Why does the basket mesh clog during a test?
Sticky excipients, gums or a tablet that swells into a gel can block the openings, and air caught in the mesh adds to it. A blocked basket stops the medium reaching the tablet, so release looks slower than it is. Clean and dry the basket fully before each run.
Where should I take each sample from?
Take every sample from the same point, about halfway between the surface of the medium and the top of the basket and away from the vessel wall. Sampling at different points gives scattered results, because the medium is never perfectly mixed.
Should I replace the medium I withdraw?
If you take several samples, either replace each one with the same volume of fresh medium at the bath temperature, or allow for the lost volume in your calculation. Note which method you used, as it changes the percentage released.
Which speed should I set?
Use the speed given in the test method for your product. The speed controller covers 25 to 150 rpm. Very low speeds stir the medium only weakly, so small changes in set-up affect the result more.
What is coning and does it affect the basket method?
Coning is a heap of fine particles that settles in a dead zone at the bottom of the vessel and dissolves slowly. It is mainly a paddle problem. With the basket method most particles stay held and stirred, though fines that escape can still settle.
Other Tablet Release and Testing Instruments
- Tablet Dissolution Test Apparatus (paddle and basket, one vessel)
- Microprocessor Dissolution Test Apparatus
- Disintegration Test Machine
For a quote, use Add To Cart or send us a message through the contact page.