Abels Flash Point Apparatus Manufacturer,Supplier and Exporter in India
Product Code : EL-GLE-4008
Jlab Export is a leading Abels Flash Point Apparatus Manufacturer,and suppliers in India
The Abel apparatus finds the closed-cup flash point of petroleum products with flash points below 70 °C. The sample warms in a covered oil cup set in a water bath. A small test flame is applied at intervals until the vapour above the oil flashes. This model has an electric heater at the bottom, controlled through an energy regulator, and works on 220 V AC.
Specifications
| Test | Closed cup flash point |
| Suitable for | Petroleum and petroleum products with flash points below 70 °C |
| Supplied with | Oil cup, cover fitted with stirrer, thermometer socket, water bath, stand and one thermometer |
| Heating | Electric heater fitted at the bottom, 220 V AC |
| Temperature control | Energy regulator |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 7–10 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Fuels and Lubricants Lab Experiments
Mechanical and chemical engineering students find the flash point of kerosene and similar light fuels on the Abel cup. They compare it with the Flash Point Pensky Martens Closed tester, used for heavier oils with higher flash points. Then they discuss why each product needs its own method.
The same lab session often covers the viscosity of a lubricating oil on a Redwood Viscometer at several temperatures. Together the two tests show how a fuel or oil behaves in storage and in use.
Frequently Asked Questions
Why must the sample be warmed slowly?
The flash point is the lowest temperature at which the vapour over the oil ignites. If the oil heats too fast, the thermometer and the vapour fall out of step and the result is wrong. Keep the heating steady with the energy regulator and stir between tests.
How is the test flame applied?
At each test temperature, stop stirring and dip the small flame into the cup through the opening in the cover, then withdraw it at once. Repeat at regular steps. The flash point is the reading when a distinct flash spreads across the vapour.
Why does a closed cup give a lower flash point than an open cup?
The cover holds the vapour over the oil, so it reaches an ignitable mixture at a lower temperature. In an open cup the vapour drifts away, and the oil must be hotter before a flash appears. Always state which method gave the figure.
Why should the sample be kept cool and sealed before the test?
Light fractions escape whenever the container is warm or open, and losing them raises the measured flash point. Store samples in closed, well-filled containers, open them just before filling the cup and use a fresh portion for each repeat.
Why should the test be kept away from draughts?
A draught blows the vapour out of the opening in the cover and disturbs the test flame, so results come out high and scattered. Work in still air, away from fans and open windows.