Digital Electrophoresis Power Supply Manufacturer,Supplier and Exporter in India

Digital Electrophoresis Power Supply

Product Code : EL-GLE-4004
Jlab Export is a leading Digital Electrophoresis Power Supply Manufacturer,and suppliers in India

An electrophoresis power supply drives current through the gel tank and holds either a set voltage or a set current through the run. This digital unit gives 0–500 V at up to 100 mA in constant voltage mode, or 1–100 mA at up to 500 V in constant current mode. Separate digital meters show the voltage and the current.

Specifications

Constant voltage mode0–500 V (100 mA max.)
Constant current mode1–100 mA (500 V max.)
ControlRange control potentiometer over the full range
MetersTwo digital meters, for voltage and current
Line and load regulation0.5 %
RippleLess than 10 mV
ProtectionShort circuit and overload protected
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Biotechnology Lab Practicals

Students connect the Horizontal Electrophoresis System red lead to red and black to black, set a moderate voltage and check that the current meter moves before they walk away. Bubbles rising from both electrodes show that the circuit is complete.

A useful class exercise records the voltage and current every few minutes, timed with a Digital Stop Watch. Students see the current climb as the buffer warms and relate the readings to Ohm’s law. Teachers can then run a Vertical Electrophoresis System in constant current mode to compare the two ways of controlling a run.

Frequently Asked Questions

Should DNA gels run at constant voltage or constant current?
Agarose DNA gels are normally run at constant voltage. The voltage sets how fast the fragments move, so runs repeat well from day to day. Constant current is used more for vertical protein gels. This unit offers both modes.

Why does the current change during a constant-voltage run?
As the buffer warms, its resistance falls, so at a fixed voltage the current rises. Buffer made too strong, or too much buffer over the gel, also raises the current. Watch the meter and lower the voltage if it keeps climbing.

Why does the current meter read zero when the voltage is set?
The circuit is open. Check that both leads are pushed fully into the supply and the tank, that the electrodes are covered by buffer and that the lid connectors are seated. A broken electrode wire gives the same reading.

When should the leads be connected or removed?
Only with the output switched off. Turn the supply off before opening the tank lid, adding buffer or lifting out the gel. Never touch the leads or the buffer while the output is on. Check leads and plugs for cracks before each run.

What do the regulation and ripple figures mean?
Line and load regulation of 0.5 % means the output stays within that margin when the mains voltage or the gel load changes. Ripple below 10 mV means the DC output is smooth, with very little AC left on it.