Oven Vacuum

Product Code : EL-RE-2068
Jlab Export is a leading Oven Vacuum Manufacturer,and suppliers in India

A vacuum oven dries samples at a lower temperature than an ordinary oven, because water and solvents boil off sooner under reduced pressure. This one has a round stainless steel chamber inside a double-walled mild steel body and works from 50 to 130 °C. The body withstands a vacuum of 760 mm Hg, and the oven is supplied without a vacuum pump.

Specifications

ConstructionDouble wall
Inner chamberRound, stainless steel
ExteriorMild steel, white stoving enamel or powder-coated paint, matt finish
Temperature range50 to 130 °C
VacuumBody withstands a vacuum of 760 mm Hg
Supply230 V AC (nominal), 50 Hz, single phase
Vacuum pumpNot included
Minimum order quantity1 unit
Dispatch timeUsually 14–18 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Pharmacy and Chemistry Lab Practicals

Pharmacy students dry heat-sensitive drug powders and wet granules at a low temperature, then work out the loss on drying. The sample is weighed in one of the tared Weighing Bottles, dried under vacuum, cooled in a Desiccator Vacuum and weighed again until two weighings agree.

Chemistry students use it to dry a recrystallised organic product that is still wet with solvent, without melting it. The oven needs a separate pump, for example the Vacuum Pump. A side-by-side run in the Hot Air Universal Oven at the same temperature shows how much faster drying goes under vacuum.

Frequently Asked Questions

Why dry under vacuum instead of in a hot air oven?
Lowering the pressure lowers the boiling point of water and solvents, so they leave the sample at a lower temperature. That suits heat-sensitive substances that would melt, decompose or darken in a hot air oven. With most of the air removed, there is also less oxygen to attack the sample.

Is a vacuum pump included?
No. The oven is supplied without a vacuum pump, so connect a lab pump of your choice. When drying solvent-wet samples, put a cold trap between the oven and the pump so the vapour does not reach the pump.

What temperature should I set?
The oven covers 50 to 130 °C. Set a temperature well below the melting or decomposition point of the sample. Under vacuum a moderate setting is usually enough, since the lower pressure does part of the work.

How do I open the oven safely after a run?
Switch off the heating, stop pumping, then let air back into the chamber slowly before you open the door. A sudden rush of air can scatter fine powders across the chamber.

What gives wrong loss-on-drying results?
Weighing the sample while it is still warm, letting it pick up moisture from room air, spreading it too thick, and stopping before two weighings agree. Cool every sample in a desiccator and weigh it promptly.