Jlab Export is a leading Vertical Autoclave Manufacturer,and suppliers in India
Saturated steam under pressure is the sterilising agent in this vertical autoclave, the upright type that is loaded through its lid. Microbiology, pathology and pharmacy labs use it for culture media, glassware, instruments and used cultures. Steam gives up its heat as it condenses on the cooler load, so it heats a load much faster than hot dry air.
Specifications
| Type | Vertical, loaded from the top |
| Sterilising agent | Saturated steam under pressure |
| Minimum order quantity | 1 unit |
| Dispatch time | Usually 14–18 working days |
| Manufacturer | JLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India |
Plate Count and Media Practicals
A viable plate count is a standard college microbiology exercise. Students make nutrient agar and tubes of saline for the dilutions, plug them with cotton wool and sterilise them together in the autoclave. Polypropylene Petri Dish (Polypropylene) plates are made to be autoclaved, so they can be sterilised and used again.
Once the agar has cooled until the flask can be held in the hand, it is poured inside a Vertical Laminar Air Flow Cabinet. The plates are spread with diluted sample and incubated, and the colonies are counted on a Microprocessor Colony Counter. At the end, every used plate and tube goes back into the autoclave before it is emptied or washed.
How Steam Sterilises the Load
Water in the chamber is boiled, and the steam drives the air out through the open vent. With the vent closed, pressure builds and the boiling point rises, so the steam gets hotter than water in an open pan ever can. It condenses on every surface it reaches and gives up its latent heat. Moist heat of this kind kills bacterial spores that survive ordinary boiling, so the load must be loose enough for steam to reach every item.
Frequently Asked Questions
Can disposable polystyrene Petri dishes go in the autoclave?
Only as waste. Polystyrene softens and melts in steam, so these plates cannot be sterilised for reuse. Put used ones in an autoclavable bag to make them safe before disposal. Glass and polypropylene dishes can be sterilised and used again.
Why must screw caps be loosened before a run?
Steam has to get into a bottle and air has to get out. A tightly closed bottle stays unsterile inside, and the pressure difference can crack the glass. Loosen each cap a little and tighten it once the load has cooled.
Does autoclave indicator tape show that the load is sterile?
No. The stripes change when the outside of a pack has met steam heat, which shows the item went through a cycle. It tells you nothing about the centre of a dense pack. For that, labs place a chemical or biological indicator inside the load.
What should be checked before every run?
Check the water level first, because the chamber must never boil dry. See that the lid seal is clean and seated, and that the vent is clear. Make sure nothing in the load touches the chamber wall or blocks the steam.
How should media and glassware be packed for a run?
Pack the chamber loosely so steam can move between items. Fill flasks of media only part way and close them with cotton wool plugs or loose caps. Wrap pipettes and glass dishes in paper or put them in cans, and keep heavy flasks at the bottom.