Digital Photo Fluorometer Manufacturer,Supplier and Exporter in India

Digital Photo Fluorometer

Product Code : EL-RE-5112
Jlab Export is a leading Digital Photo Fluorometer Manufacturer,and suppliers in India

A photodiode detector in this digital photo fluorometer measures the light a sample gives off when a 12 V, 50 W tungsten halogen lamp excites it. That emitted light has a longer wavelength than the exciting light, and filters keep the two apart. Samples sit in 15 x 80 mm test tubes, with at least 4 ml for a reading.

Specifications

Fluoro sensitivityFull-scale deflection with 1 ppm quinine sulphate in 0.1 N sulphuric acid (primary wavelength 360 nm)
Display3 digit, bright red 7-segment LED
Excitation source12 V, 50 W tungsten halogen lamp
SensitivityAdjustable in 4 ranges
DetectorWide-range photodiode/cell
Sample test tube15 mm (D) x 80 mm (H)
Minimum sample4 ml
Power230 V AC ±10%, 50 Hz
Size260 x 335 x 160 mm (approx.)
Minimum order quantity1 unit
Dispatch timeUsually 7–10 working days
ManufacturerJLab (Jain Laboratory Instruments Pvt Ltd), Ambala, India

Fluorescence Experiments for Pharmacy Students

In pharmaceutical analysis practicals, students make a quinine sulphate stock in dilute sulphuric acid and dilute it into a series in Volumetric Flasks, Class A Borosilicate Glass Plastic Stopper JLab flasks. They plot fluorescence against concentration and use the line to find the quinine in tonic water or a tablet extract.

Riboflavin from a vitamin tablet is estimated the same way against a riboflavin standard, with volumes taken by Graduated Pipettes, Class B Soda Glass JLab. Teachers can also read the riboflavin series on a Digital Photo Colorimeter. The fluorometer still gives a clear signal at dilutions where the absorbance is too small to read.

Frequently Asked Questions

What does full-scale deflection with 1 ppm quinine sulphate mean?
It is the sensitivity figure. A 1 ppm solution of quinine sulphate in 0.1 N sulphuric acid, excited at the 360 nm primary wavelength, brings the display to full scale. Weaker solutions read lower on the same setting.

Do I need a fluorometer or a colorimeter?
A colorimeter measures how much light a coloured solution absorbs. A fluorometer measures the light a substance gives out at a longer wavelength after it is excited, so it works only for fluorescent compounds but reads far weaker solutions of them.

Which of the 4 sensitivity ranges should I start on?
Start on the least sensitive range with your strongest standard, then step up until that standard reads near the top of the display without going over. Keep that range for the blank, the standards and the samples.

Why should riboflavin solutions be kept out of strong light?
Riboflavin breaks down in light, so a standard left on a sunny bench gives a falling reading. Make the solutions fresh, keep the flasks covered or in a dark cupboard, and read the standards and samples in one session.

Why does the reading drift just after switching on?
The tungsten halogen lamp and the detector take time to warm up and settle. Switch on early, then set the blank and the standard again just before you read the samples.