• +91-85699-09696

Venturi meter and Orifice Meter Apparatus Manufacturer,Supplier and Exporter in India

Venturi meter and Orifice Meter Apparatus

Product Code : EL-HLE-8958
Jlab Export is a leading Venturi meter and Orifice Meter Apparatus Manufacturer,and suppliers in India

Venturi meter and Orifice Meter Apparatus (code EL-HLE-8958) is a pipe-flow rig for calibrating a 25 mm venturi meter and a 25 mm orifice meter, set in two GI pipe lines. A 0.5 HP single-phase pump draws water from a sump, and distribution chambers with isolating cocks allow each meter to be tested on its own. JLab makes this venturi and orifice meter rig in Ambala, India.

Specifications

ParameterDetail
Pipe lines2, GI, each with a gate valve
MetersVenturi meter and orifice meter, 25 mm size
Meter selectionDistribution chambers with isolating cocks, for testing one meter at a time
Pump0.5 HP, single phase, from the sump, with control valve
Power supply requiredSingle phase, 200/220 V
Water requiredTo fill the sump tank
Quality checkEach flow meter apparatus is visually inspected before dispatch
Pack quantityEach (sold individually)
Minimum order quantity1 unit
Dispatch timeUsually 14–18 working days
Product codeEL-HLE-8958
ManufacturerJain Laboratory Instruments Pvt Ltd (JLab) — OEM
Country of originIndia (made in Ambala, Haryana)

Calibrating the Two Meters

Both meters narrow the flow and measure the pressure drop it causes. With inlet area a1, throat or orifice area a2 and differential head h, the theoretical discharge is Q = a1 a2 √(2gh) / √(a1² − a2²). The actual discharge divided by this figure is the meter’s coefficient of discharge Cd, and repeating the test at several gate valve settings gives a calibration curve of Q against √h for each meter.

Venturi Against Orifice

The venturi meter narrows gradually and opens again through a long diverging cone, so most of the pressure drop is recovered downstream and its Cd lies close to 1. The orifice plate contracts the flow abruptly into a vena contracta and leaves a turbulent wake behind it, so its Cd is much lower and its permanent head loss far higher. The orifice plate is simpler and much shorter to fit in a pipe line, which is why both types remain in use.

Other Flow Meter and Discharge Rigs

The rig itself belongs to Hydrology Lab Equipments, a part of Mechanical Engineering Lab Equipment.

Request a Quote

Use Add To Cart or the contact page to ask for a quotation, stating the quantity and the delivery town.

Last updated: 30 October 2026


   

Office Address

Works: 2475-84,
Hargolal Road,Ambala, Haryana

[email protected]

+91-85699-09696

Follow Us